CLASS I ONGOING Device recall · Reported 16 Jul 2025 · FDA Enforcement Report

Hillrom LikoScale component recalled by Baxter Healthcare

Baxter Healthcare is recalling Hillrom LikoScale component. The recall was initiated on 30 May 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesAll LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025
Quantity recalled9839 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2032-2025
FDA event ID
96856
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
30 May 2025
FDA classified
9 Jul 2025
Reported
16 Jul 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

30 May 2025Recall initiated by company
9 Jul 2025Classified by FDA+40 days
16 Jul 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Product as listed by the FDA

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Where it was distributed

Worldwide

Questions about this recall

Why was it recalled?

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Which lots are affected?

All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 96856
ReportedClassRecallCompanyReasonWhere
CLASS IHillrom Q-Link 13 component recalled by Baxter HealthcareBaxter HealthcareOtherIL

More recalls from Baxter Healthcare

All 577

Other Wheelchairs, Beds & Lifts recalls

All 174