CLASS II Device recall · Reported 15 Oct 2014 · FDA Enforcement Report

The Hulka Clip is a sterile packaged medical device recalled

Richard Wolf Medical Instruments is recalling The Hulka Clip is a sterile packaged medical device that carries a two year, distributed nationwide in the US. The recall was initiated on 9 Jul 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code: 4986.09; Lot: 51000232
Quantity recalled73 cartons (1,460 clips)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0037-2015
FDA event ID
68777
Recalling firm
Richard Wolf Medical Instruments, Vernon Hills, IL
Classification
Class II
Status
Terminated
Recall initiated
9 Jul 2014
FDA classified
7 Oct 2014
Reported
15 Oct 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

9 Jul 2014Recall initiated by company
7 Oct 2014Classified by FDA+90 days
15 Oct 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed final release testing, but the subsequent lot failed biocompatibility testing on a component that was also used to manufacture lot 51000232.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

The Hulka Clip is a sterile packaged medical device that carries a two year expiration date. It is packaged two clips per procedural tray. Theses procedural trays are then sealed in a Tyvek/Mylar pouch and packaged in a cardboard shipping carton with ten pouches in each shipping carton.

Where it was distributed

Worldwide distribution: US (Nationwide) including states of: (AK, AL, CO, CT, FL, GA, IA, IN, KY, LA, MI, MO, MS, NC, NY, OH, OR, PA, SC, TN, TX, VA, WA, WV; and country of: Germany.

Nationwide AlaskaAlabamaColoradoConnecticutFloridaGeorgiaIowaIndianaKentuckyLouisianaMichiganMissouriMississippiNorth CarolinaNew YorkOhioOregonPennsylvaniaSouth CarolinaTennessee

Questions about this recall

Why was it recalled?

Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed final release testing, but the subsequent lot failed biocompatibility testing on a component that was also used to manufacture lot 51000232.

Which lots are affected?

Product Code: 4986.09; Lot: 51000232

Where was it sold?

Worldwide distribution: US (Nationwide) including states of: (AK, AL, CO, CT, FL, GA, IA, IN, KY, LA, MI, MO, MS, NC, NY, OH, OR, PA, SC, TN, TX, VA, WA, WV; and country of: Germany.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Richard Wolf Medical Instruments

All 9

Other Hip & Knee Implants recalls

All 3,097