CLASS II Device recall · Reported 17 Apr 2013 · FDA Enforcement Report

Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD recalled

Siemens Medical Solutions USA is recalling Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, distributed nationwide in the US. The recall was initiated on 15 Jan 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number 5902767 with serial numbers 2217, 2038, 1228, 1007, 1219, 2156, 2264, 2190, 2132, 2213, 1484, 2166, 2259, 2068, 1610, 1601, 1413, 2236, 2153, 1444, 1477, 1632, 2037, 2180, 1663, and 1612; Model Number 5902783 with serial numbers 1003, 1401, 1402, and 1405; Model Number 7152478 with serial numbers 2580, 2524, 2590, 2536, 2563, and 2534; Model Number 10093962 with serial numbers 5837, 5120, 5523, 5910, 5507, 6141, 4112, 6274, 4506, 5900, 4121, 6245, 6242, 4509, and 6113.
Quantity recalled51

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1097-2013
FDA event ID
64678
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Terminated
Recall initiated
15 Jan 2013
FDA classified
11 Apr 2013
Reported
17 Apr 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

15 Jan 2013Recall initiated by company
11 Apr 2013Classified by FDA+86 days
17 Apr 2013Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is possible in cases where the table vertical lift movement is obstructed by a solid object which could potentially result in collision. This can re

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. Angiographic x-ray system.

Where it was distributed

Nationwide Distribution including the states of CA, FL, GA, IA, IN, KS, LA, ME, MI, MO, NC, NH, NJ, NY, NV, OH, PA, TN, TX, VA and WA.

Nationwide CaliforniaFloridaGeorgiaIowaIndianaKansasLouisianaMaineMichiganMissouriNorth CarolinaNew HampshireNew JerseyNevadaNew YorkOhioPennsylvaniaTennesseeTexasVirginia

Questions about this recall

Is Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic recalled?

Yes. Siemens Medical Solutions USA recalled Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic on 15 Jan 2013. The recall is Class II and its status is Terminated. Recall number Z-1097-2013.

Why was it recalled?

Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is possible in cases where the table vertical lift movement is obstructed by a solid object which could potentially result in collision. This can re

Which lots are affected?

Model Number 5902767 with serial numbers 2217, 2038, 1228, 1007, 1219, 2156, 2264, 2190, 2132, 2213, 1484, 2166, 2259, 2068, 1610, 1601, 1413, 2236, 2153, 1444, 1477, 1632, 2037, 2180, 1663, and 1612; Model Number 5902783 with serial numbers 1003, 1401, 1402, and 1405; Model Number 7152478 with serial numbers 2580, 2524, 2590, 2536, 2563, and 2534; Model Number 10093962 with serial numbers 5837, 5120, 5523, 5910, 5507, 6141, 4112, 6274, 4506, 5900, 4121, 6245, 6242, 4509, and 6113.

Where was it sold?

Nationwide Distribution including the states of CA, FL, GA, IA, IN, KS, LA, ME, MI, MO, NC, NH, NJ, NY, NV, OH, PA, TN, TX, VA and WA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.