CLASS II Device recall · Reported 18 Mar 2020 · FDA Enforcement Report

Iliac Screws, multiple sizes, Article Nos recalled

Medicrea International is recalling Iliac Screws, multiple sizes, Article Nos, distributed nationwide in the US. The recall was initiated on 28 Jan 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLots 19D0683 19D0684 19D0685 19D0686 19D0369 19H0470 19D0370 19H0473 19D0371 19G1071 19L0501 19D0372 19G1073 19D0570 19G1075 19J0445 19D0571 19G1076 19J0867 19D0572 19G0521 19J0868 19L0526 19L0502 19L0525 19D0573 19G0522 19J0869 19D0574 19G1077 19L0503 19D0575 19H0474 19D0576 19G0523 19G1078 19D0577 19G1079 19H0475
Quantity recalled3529

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1473-2020
FDA event ID
84928
Recalling firm
Medicrea International, Rillieux La Pape
Classification
Class II
Status
Terminated
Recall initiated
28 Jan 2020
FDA classified
11 Mar 2020
Reported
18 Mar 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

28 Jan 2020Recall initiated by company
11 Mar 2020Classified by FDA+43 days
18 Mar 2020Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Product as listed by the FDA

Iliac Screws, multiple sizes, Article Nos. B36020500 B36020570 B36020580 B36020590 B36027500 B36027570 B36027580 B36027590 B36028500 B36028570 B36028580 B36028590 B36029500 B36029570 B36029580 B36029590 Screws manufactured in titanium alloy Ti-6Al-4V LI conforming to ISO5832-3 and ASTM. F136 specifications. These screws are part of an internal fixation system composed of hooks, rods, plates, cross links, connection and locking devices.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.

Nationwide ColoradoFloridaGeorgiaLouisianaOhioRhode Island

Questions about this recall

Why was it recalled?

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Which lots are affected?

Lots 19D0683 19D0684 19D0685 19D0686 19D0369 19H0470 19D0370 19H0473 19D0371 19G1071 19L0501 19D0372 19G1073 19D0570 19G1075 19J0445 19D0571 19G1076 19J0867 19D0572 19G0521 19J0868 19L0526 19L0502 19L0525 19D0573 19G0522 19J0869 19D0574 19G1077 19L0503 19D0575 19H0474 19D0576 19G0523 19G1078 19D0577 19G1079 19H0475

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 84928

This product is part of a recall event; the main page for the event is Pedicle Screws, multiple sizes, Article Nos recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIPedicle Screws, multiple sizes, Article Nos recalledMedicrea InternationalOtherNationwide

More recalls from Medicrea International

All 22

Other Other Medical Devices recalls

All 13,969