CLASS II ONGOING Device recall · Reported 15 May 2024 · FDA Enforcement Report
Intelli-C: Right side suspended Tabletop recalled by Nrt X-ray A/s
Nrt X-ray A/s is recalling Intelli-C: Right side suspended Tabletop, distributed nationwide in the US. The recall was initiated on 17 Apr 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1784-2024
- FDA event ID
- 94363
- Recalling firm
- Nrt X-ray A/s, Hasselager, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 17 Apr 2024
- FDA classified
- 7 May 2024
- Reported
- 15 May 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Imaging Equipment
Timeline
| 17 Apr 2024 | Recall initiated by company | |
| 7 May 2024 | Classified by FDA | +20 days |
| 15 May 2024 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
Where it was distributed
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
Nationwide HawaiiIowaKentuckyLouisianaMassachusettsMichiganMinnesotaNorth CarolinaNevadaNew YorkPennsylvaniaSouth CarolinaWashingtonWisconsinWest Virginia
Questions about this recall
Is Intelli-C: Right side suspended Tabletop recalled?
Yes. Nrt X-ray A/s recalled Intelli-C: Right side suspended Tabletop on 17 Apr 2024. The recall is Class II and its status is Ongoing. Recall number Z-1784-2024.
Why was it recalled?
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
Which lots are affected?
REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023
Where was it sold?
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nrt X-ray A/s
All 4| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Intelli-C, Right side suspended Tabletop recalled by Nrt X-ray A/s | Nrt X-ray A/s | Device Malfunction | Nationwide |
| Device | CLASS II | Intelli-C, Left side suspended Tabletop recalled by Nrt X-ray A/s | Nrt X-ray A/s | Device Malfunction | Nationwide |
| Device | CLASS II | Footrest ( use exclusively in combination recalled by Nrt X-ray A/s | Nrt X-ray A/s | Foreign Material | 10 states |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device is recalled | Canon Medical System, USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederland B.v | Lead & Heavy Metals |