CLASS II Device recall · Reported 24 Dec 2014 · FDA Enforcement Report
Intended for both adult and pediatric patients recalled
Philips Medical Systems is recalling Intended for both adult and pediatric patients for taking diagnostic radiographic, distributed nationwide in the US. The recall was initiated on 8 Apr 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0508-2015
- FDA event ID
- 67869
- Recalling firm
- Philips Medical Systems, Andover, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 8 Apr 2014
- FDA classified
- 16 Dec 2014
- Reported
- 24 Dec 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 8 Apr 2014 | Recall initiated by company | |
| 16 Dec 2014 | Classified by FDA | +252 days |
| 24 Dec 2014 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.
Product as listed by the FDA
Intended for both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with the patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.
Where it was distributed
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI.,and the countries of : Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iraq, Ireland, Italy, Jordan, Liberia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Arab Emirates and United Kingdom.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan
Questions about this recall
Why was it recalled?
In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.
Which lots are affected?
Serial Numbers: SN12000024, SN12000166 , SN13000081 , SN13000127 , SN13000140 , SN13000141 , SN13000143 , SN13000144 , SN13000145 , SN13000167 , SN13000168 , SN13000169 , SN13000196 , SN13000199 , SN13000200 , SN13000201 , SN13000202 , SN13000203 , SN13000204 , SN13000205 , SN14000002 , SN14000004 , SN14000005 , SN14000006 , SN14000008 , SN14000009 , SN14000010 , SN14000011 , SN14000013 , SN14000014 , SN14000015 , SN14000016 , SN14000017 , SN14000019 , SN14000022 , SN14000024 , SN14000026 , SN14000028 , SN14000030 , SN14000031 , SN14000032 , SN14000033 , SN14000038 , SN14000041 , SN14000046 , SN14000047 , SN14000048 , SN13000153 , SN13000171 , SN13000155 , SN13000165 , SN13000187 , SN13000170 , SN13000176 , SN13000139 , SN13000189 , SN7369686 , SN13000150 , SN13000186 , SN13000197 , SN13000152 , SN13000162 , SN13000163 , SN14000012 , SN11000018 , SN13000136 , SN13000137 , SN13000138 , SN13000147 , SN13000154 , SN13000156 , SN13000159 , SN13000160 , SN13000164 , SN13000172 , SN13000175 , SN13000181 , SN13000188 , SN13000191 , SN13000194 , SN14000027 , SN130000184 , SN13000132 , SN13000146 , SN13000148 , SN13000149 , SN13000151 , SN13000161 , SN13000174 , SN13000177 , SN13000179 , SN13000179 , SN13000180 , SN13000183 , SN13000185 , SN13000190 , SN13000193 , SN13000206 , SN14000003 , SN14000023 , SN14000036 , SN11000001 , SN11000002 , SN11000003 , SN11000004 , SN11000005 , SN11000006 , SN11000007 , SN11000008 , SN11000009 , SN11000010 , SN11000011 , SN11000012 , SN11000013 , SN11000014 , SN11000014 , SN11000015 , SN11000016 , SN11000017 , SN11000020 , SN11000021 , SN11000021 , SN11000022 , SN11000023 , SN11000024 , SN11000025 , SN11000026 , SN11000027 , SN11000028 , SN11000029 , SN11000030 , SN11000031 , SN11000032 , SN11000033 , SN11000034 , SN11000035 , SN11000036 , SN11000037 , SN11000039 , SN11000040 , SN11000041 , SN11000042 , SN11000044 , SN11000045 , SN11000046 , SN11000047 , SN11000048 , SN11000048 , SN11000049 , SN11000050 , SN11000051 , SN11000052 , SN11000054 , SN11000055 , SN11000056 , SN11000057 , SN11000058 , SN11000059 , SN11000060 , SN11000061 , SN11000062 , SN11000063 , SN11000064 , SN11000065 , SN11000066 , SN11000067 , SN11000068 , SN11000069 , SN11000070 , SN11000071 , SN11000072 , SN11000073 , SN11000074 , SN11000075 , SN11000076 , SN11000077 , SN11000078 , SN11000079 , SN11000080 , SN11000081 , SN11000082 , SN11000083 , SN11000084 , SN12000002 , SN12000003 , SN12000004 , SN12000005 , SN12000006 , SN12000007 , SN12000008 , SN12000009 , SN12000010 , SN12000012 , SN12000013 , SN12000014 , SN12000015 , SN12000016 , SN12000018 , SN12000019 , SN12000020 , SN12000021 , SN12000022 , SN12000023 , SN12000025 , SN12000026 , SN12000027 , SN12000028 , SN12000029 , SN12000030 , SN12000031 , SN12000032 , SN12000033 , SN12000034 , SN12000035 , SN12000036 , SN12000037 , SN12000038 , SN12000039 , SN12000040 , SN12000041 , SN12000042 , SN12000043 , SN12000044 , SN12000045 , SN12000046 , SN12000047 , SN12000048 , SN12000049 , SN12000050 , SN12000051 , SN12000052 , SN12000053 , SN12000054 , SN12000055 , SN12000056 , SN12000057 , SN12000058 , SN12000059 , SN12000060 , SN12000061 , SN12000062 , SN12000063 , SN12000064 , SN12000065 , SN12000066 , SN12000067 , SN12000068 , SN12000069 , SN12000070 , SN12000071 , SN12000072 , SN12000073 , SN12000074 , SN12000075 , SN12000076 , SN12000077 , SN12000078 , SN12000079 , SN12000080 , SN12000081 , SN12000082 , SN12000083 , SN12000083 , SN12000084 , SN12000085 , SN12000086 , SN12000087 , SN12000088 , SN12000089 , SN12000090 , SN12000091 , SN12000092 , SN12000093 , SN12000094 , SN12000095 , SN12000098 , SN12000099 , SN12000100 , SN12000101 , SN12000102 , SN12000103 , SN12000104 , SN12000105 , SN12000106 , SN12000107 , SN12000108 , SN12000109 , SN12000110 , SN12000111 , SN12000112 , SN12000113 , SN12000114 , SN12000115 , SN12000116 , SN12000117 , SN12000118 , SN12000119 , SN12000120 , SN12000121 , SN12000122 , SN12000123 , SN12000124 , SN12000125 , SN13000001 , SN13000002 , SN13000004 , SN13000005 , SN13000006 , SN13000007 , SN13000008 , SN13000010 , SN13000011 , SN13000012 , SN13000013 , SN13000014 , SN13000015 , SN13000016 , SN13000017 , SN13000018 , SN13000019 , SN13000020 , SN13000021 , SN13000022 , SN13000023 , SN13000025 , SN13000026 , SN13000027 , SN13000028 , SN13000029 , SN13000030 , SN13000031 , SN13000032 , SN13000033 , SN13000034 , SN13000035 , SN13000036 , SN13000037 , SN13000038 , SN13000039 , SN13000040 , SN13000041 , SN13000042 , SN13000043 , SN13000044 , SN13000045 , SN13000046 , SN13000047 , SN13000048 , SN13000049 , SN13000050 , SN13000051 , SN13000052 , SN13000053 , SN13000054 , SN13000055 , SN13000056 , SN13000057 , SN13000058 , SN13000059 , SN13000060 , SN13000061 , SN13000062 , SN13000063 , SN13000064 , SN13000065 , SN13000066 , SN13000067 , SN13000069 , SN13000070 , SN13000071 , SN13000072 , SN13000073 , SN13000074 , SN13000075 , SN13000076 , SN13000077 , SN13000078 , SN13000079 , SN13000080 , SN13000082 , SN13000083 , SN13000084 , SN13000085 , SN13000086 , SN13000087 , SN13000088 , SN13000089 , SN13000090 , SN13000091 , SN13000092 , SN13000093 , SN13000094 , SN13000095 , SN13000096 , SN13000097 , SN13000098 , SN13000099 , SN13000100 , SN13000101 , SN13000102 , SN13000103 , SN13000104 , SN13000105 , SN13000106 , SN13000107 , SN13000108 , SN13000109 , SN13000110 , SN13000111 , SN13000112 , SN13000113 , SN13000114 , SN13000115 , SN13000116 , SN13000117 , SN13000118 , SN13000119 , SN13000120 , SN13000121 , SN13000122 , SN13000123 , SN13000124 , SN13000124 , SN13000125 , SN13000126 , SN13000128 , SN13000129 , SN13000130 , SN13000131 , SN13000133 , SN13000134 , SN13000135 , SN13000157 , SN13000173 , SN13000178 , SN13000182 , SN13000192 , SN13000198 , SN14000007 , SN14000018 , SN14000021 , SN14000025. Four systems were delivered to foreign (non-US) countries that do not have serial numbers.
Where was it sold?
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI.,and the countries of : Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iraq, Ireland, Italy, Jordan, Liberia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Arab Emirates and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Medical Systems
All 81| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | IQon Spectral CT IQon Spectral CT Upgrades recalled | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Philips Spectral CT on Rails recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Model: Incisive CT; Product Code (REF): (1) 728143, (2) recalled | Philips Medical Systems | Device Malfunction | 18 states |
Other Spinal Implants recalls
All 1,157| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |