CLASS II Device recall · Reported 19 Sep 2012 · FDA Enforcement Report
Interfaces subject to recall: Medtronic NIM recalled by Medtronic Xomed
Medtronic Xomed is recalling Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient, distributed nationwide in the US. The recall was initiated on 27 Sep 2010 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2344-2012
- FDA event ID
- 63008
- Recalling firm
- Medtronic Xomed, Jacksonville, FL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 27 Sep 2010
- FDA classified
- 10 Sep 2012
- Reported
- 19 Sep 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Spinal Implants
Timeline
| 27 Sep 2010 | Recall initiated by company | |
| 10 Sep 2012 | Classified by FDA | +714 days |
| 19 Sep 2012 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253210 (Pulse). LABELING: Box Label: REF 8253210 x1***NIM-Response 3.0 Patient Interface***Medtronic Xomed, Inc., 6743 Southpoint Drive North Jacksonville, FL 32216-0980 USA***Rx Only***Lot***SN:***" Top of unit is labeled in part: "***NIM-Response 3.0***Medtronic***" Bottom/Underside of unit is labeled in part: "***5X20mm FA.L., 32mA. 250VAC***REF***MEDTRONIC XOMED INC. JACKSONVILLE, FL USA***" Cord is labeled in part: "***Ref.***Mfg Date***Sn***" The NIM 3.0 Patient Interfaces [REF #'s 8253210 (Pulse), 8253200 (Response), 8253410 (Neuro)] is an EMG Nerve Integrity Monitor intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots.
Where it was distributed
Worldwide distribution: USA (nationwide) and countries of: Australia, Austria, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Iran, Ireland, Israel, Italy, Japan, Lebanon, Libya, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudia Arabia, Singapore, Slovak, Spain, SRI Lanka, Sweden, Turkey, United Arab Emirates, and United Kingdom.
Questions about this recall
Why was it recalled?
In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
Which lots are affected?
8253210 (Pulse). Serial Numbers 00004, 00009, 00010, 00013, 00014, 00015, 00017 (International Consignees only)
Where was it sold?
Worldwide distribution: USA (nationwide) and countries of: Australia, Austria, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Iran, Ireland, Israel, Italy, Japan, Lebanon, Libya, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudia Arabia, Singapore, Slovak, Spain, SRI Lanka, Sweden, Turkey, United Arab Emirates, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 63008This product is part of a recall event; the main page for the event is Interfaces subject to recall: Medtronic NIM recalled by Medtronic Xomed.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Interfaces subject to recall: Medtronic NIM recalled by Medtronic Xomed | Medtronic Xomed | Device Malfunction | Nationwide | |
| CLASS II | Interfaces subject to recall: Medtronic NIM recalled by Medtronic Xomed | Medtronic Xomed | Device Malfunction | Nationwide |
More recalls from Medtronic Xomed
All 25| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | IPC Powerease System recalled by Medtronic Xomed | Medtronic Xomed | Other | 25 states |
| Device | CLASS II | NIM Vital Nerve Monitoring System: Console NIM4CM01 NIM recalled | Medtronic Xomed | Device Malfunction | Nationwide |
| Device | CLASS II | Medtronic ENT Ultra Round Steel Cutting Bur recalled by Medtronic Xomed | Medtronic Xomed | Other | FL |
| Device | CLASS II | NIM TriVantage EMG Endotracheal Tubes recalled by Medtronic Xomed | Medtronic Xomed | Other | FL |
| Device | CLASS I | NIM EMG Endotracheal Tube, Part Numbers: a) recalled by Medtronic Xomed | Medtronic Xomed | Other | Nationwide |
Other Spinal Implants recalls
All 1,159| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | Avid Medical | Sterility | Nationwide |