CLASS II ONGOING Device recall · Reported 28 Oct 2026 · FDA Enforcement Report
Vacuette LH Lithium Heparin Separator recalled
Greiner Bio-One North America is recalling Vacuette LH Lithium Heparin Separator, distributed nationwide in the US. The recall was initiated on 29 Jul 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0200-2027
- FDA event ID
- 99518
- Recalling firm
- Greiner Bio-One North America, Monroe, NC
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 29 Jul 2026
- FDA classified
- 21 Oct 2026
- Reported
- 28 Oct 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Syringes & Needles
Timeline
| 29 Jul 2026 | Recall initiated by company | |
| 21 Oct 2026 | Classified by FDA | +84 days |
| 28 Oct 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Product as listed by the FDA
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Where it was distributed
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Nationwide ColoradoFloridaIllinoisMissouriNebraskaNew YorkOhioPennsylvaniaVirginia
Questions about this recall
Why was it recalled?
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Which lots are affected?
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
Where was it sold?
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 3 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Greiner Bio-One North America
All 17| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Vacuette CAT Serum Separator Clot Activator recalled | Greiner Bio-One North America | Potency & Assay | 28 states |
| Device | CLASS II | Greiner Bio-one Vacuette 9NC Coagulation sodium citrate recalled | Greiner Bio-One North America | Other | Nationwide |
| Device | CLASS II | Vacuette¿ Tube Cat Serum Separator Clot Activator red recalled | Greiner Bio-One North America | Device Malfunction | Nationwide |
| Device | CLASS II | Vacuette Tube FX Sodium Fluoride/Potassium Oxalate recalled | Greiner Bio-One North America | Other | Nationwide |
| Device | CLASS II | greiner bio-one Vacuette Tube K3E K3EDTA recalled | Greiner Bio-One North America | Labeling Errors | Nationwide |
Other Syringes & Needles recalls
All 1,069| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medtronic Biopsy Needle Kit recalled by Medtronic Navigation | Medtronic Navigation | Sterility | 44 states |
| Device | CLASS II | Williams Cystoscopic Injection Needles, used recalled by Cook | Cook | Device Malfunction | 5 states |
| Device | CLASS II | Embosphere 500-700 micrometer Syringe Sterile Steam recalled | Merit Medical Systems | Sterility | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |