CLASS II Device recall · Reported 16 Jan 2013 · FDA Enforcement Report

(l) Visum Halogen Surgical Light, Visum LED Surgical Light recalled

Stryker Communications is recalling (l) Visum Halogen Surgical Light, Visum LED Surgical Light, Stryker Flat Panel and, distributed nationwide in the US. The recall was initiated on 25 Sep 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCode Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG, 0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING.
Quantity recalledUnited States: 1,082 units and Outside United States: 199 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0657-2013
FDA event ID
63355
Recalling firm
Stryker Communications, Flower Mound, TX
Classification
Class II
Status
Terminated
Recall initiated
25 Sep 2012
FDA classified
9 Jan 2013
Reported
16 Jan 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

25 Sep 2012Recall initiated by company
9 Jan 2013Classified by FDA+106 days
16 Jan 2013Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e

Product as listed by the FDA

(l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the Stryker Spring Arm used in multiple configurations as a component of the (1)Visum Halogen Surgical Light, (2)Visum LED Surgical Light, and (3)Stryker Flat Panel and Navigation Arm. The Spring Arm that is subject to this field action suspends loads, is a conduit for power/data, and facilitates multi-directional movement. The Spring Arms that have been identified within scope are replacement parts that have potentially been serviced incorrectly.

Where it was distributed

Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.

Nationwide

Questions about this recall

Is (l) Visum Halogen Surgical Light, Visum LED Surgical Light, Stryker Flat Panel and recalled?

Yes. Stryker Communications recalled (l) Visum Halogen Surgical Light, Visum LED Surgical Light, Stryker Flat Panel and on 25 Sep 2012. The recall is Class II and its status is Terminated. Recall number Z-0657-2013.

Why was it recalled?

Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e

Which lots are affected?

Code Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG, 0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING.

Where was it sold?

Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Communications

All 20

Other Surgical Instruments recalls

All 3,809