CLASS III Device recall · Reported 22 May 2013 · FDA Enforcement Report
Liquichek" ToRCH Plus IgM Control recalled by Bio-rad Laboratories
Bio-rad Laboratories is recalling Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended, distributed in Minnesota, New Mexico, New York, Oregon, Puerto Rico, Wisconsin. The recall was initiated on 15 Apr 2013 and the FDA has classified it as Class III.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1330-2013
- FDA event ID
- 64959
- Recalling firm
- Bio-rad Laboratories, Irvine, CA
- Classification
- Class III
- Status
- Terminated
- Recall initiated
- 15 Apr 2013
- FDA classified
- 16 May 2013
- Reported
- 22 May 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Potency & Assay
- Device type
- Patient Monitors
Timeline
| 15 Apr 2013 | Recall initiated by company | |
| 16 May 2013 | Classified by FDA | +31 days |
| 22 May 2013 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an unassayed quality control serum to monitor precision of IgM laboratory testing procedures for the analytes listed in this package insert. This product is not intended for use in blood donor screening assays. Analytes claimed with this lot of Liquichek ToRCH Plus IgM Control: ¿ Cytomegalovirus (CMV) IgM ¿ Herpes Simplex Virus Type 1/2 (HSV-1/2) IgM ¿ Rubella Virus IgM ¿ Toxoplasma gondii IgM
Where it was distributed
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
Questions about this recall
Why was it recalled?
The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Which lots are affected?
Lot Number 18972, Catalog Number 229 and 229PX.
Where was it sold?
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class III mean?
Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Bio-rad Laboratories
All 27| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Platelia Toxo IgM recalled by Bio-Rad Laboratories | Bio-Rad Laboratories | Potency & Assay | Nationwide |
| Device | CLASS II | Bio-Rad, REF: 12005660, QXDx Bcr-abl %IS Kit recalled | Bio-Rad Laboratories | Other | Nationwide |
| Device | CLASS II | BioPlex 2200 REF 665-2050, Apls IgM Pack, Apls IgM Reagent Pack recalled | Bio-Rad Laboratories | Packaging Defects | Nationwide |
| Device | CLASS II | BioPlex 2200, Anti-CCP Reagent Pack recalled by Bio-Rad Laboratories | Bio-Rad Laboratories | Other | Nationwide |
| Device | CLASS II | BioPlex 2200, Anti-CCP Calibrator Set recalled by Bio-Rad Laboratories | Bio-Rad Laboratories | Other | Nationwide |
Other Patient Monitors recalls
All 1,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Passport Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | T1 Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | cPM Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | BeneVision Series Telemetry Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | Panorama and Telemetry Monitoring System recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |