CLASS II Device recall · Reported 16 Sep 2020 · FDA Enforcement Report
Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1 recalled
Talladium is recalling Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, distributed nationwide in the US. The recall was initiated on 6 Aug 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2943-2020
- FDA event ID
- 86227
- Recalling firm
- Talladium, Valencia, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Aug 2020
- FDA classified
- 9 Sep 2020
- Reported
- 16 Sep 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Foreign Material
- Device type
- Dental Devices
Timeline
| 6 Aug 2020 | Recall initiated by company | |
| 9 Sep 2020 | Classified by FDA | +34 days |
| 16 Sep 2020 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.
Where it was distributed
U.S. Nationwide distribution including in the states of CA, FL, KS, LA, MO, and PA. No OUS Consignees.
Nationwide CaliforniaFloridaKansasLouisianaMissouriPennsylvania
Questions about this recall
Why was it recalled?
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.
Which lots are affected?
Lot Number.: 1AZG
Where was it sold?
U.S. Nationwide distribution including in the states of CA, FL, KS, LA, MO, and PA. No OUS Consignees.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Talladium
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Luminesse ML PMMA, 98x20mm recalled by Talladium | Talladium | Labeling Errors | Nationwide |
Other Dental Devices recalls
All 426| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Grandio Universal Nano-Hybrid Restorative Material - Capsules 0.25Gm recalled | Henry Schein | Packaging Defects | Nationwide |
| Device | CLASS II | GM Helix Implant, Endosseous Dental Implant Article No. recalled | Straumann USA | Packaging Defects | Nationwide |
| Device | CLASS II | KL Owen Custom Orthodontic Solution recalled by Klowen Braces | Klowen Braces | Device Malfunction | Nationwide |
| Device | CLASS II | Straumann n!ce PMMA Full-arch Restoration recalled | Straumann USA | Other | 8 states |
| Device | CLASS II | Straumann n!ce Zr, HT, Full-arch Restoration recalled | Straumann USA | Other | 8 states |