CLASS II ONGOING Device recall · Reported 11 Feb 2026 · FDA Enforcement Report
Lumipulse G pTau 217 Plasma Immunoreaction Cartridges recalled
Fujirebio Diagnostics is recalling Lumipulse G pTau 217 Plasma Immunoreaction Cartridges, distributed nationwide in the US. The recall was initiated on 11 Dec 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1302-2026
- FDA event ID
- 98232
- Recalling firm
- Fujirebio Diagnostics, Malvern, PA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 11 Dec 2025
- FDA classified
- 5 Feb 2026
- Reported
- 11 Feb 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Labeling Errors
- Device type
- Test Kits & Diagnostics
Timeline
| 11 Dec 2025 | Recall initiated by company | |
| 5 Feb 2026 | Classified by FDA | +56 days |
| 11 Feb 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Product as listed by the FDA
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Where it was distributed
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Nationwide ArizonaCaliforniaFloridaIndianaKansasKentuckyMassachusettsMarylandMinnesotaNew JerseyPennsylvaniaTexasWashington
Questions about this recall
Why was it recalled?
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Which lots are affected?
Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;
Where was it sold?
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
6 other products · FDA event 98232This product is part of a recall event; the main page for the event is Lumipulse pTau 217 Plasma Controls recalled by Fujirebio Diagnostics.
More recalls from Fujirebio Diagnostics
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators recalled | Fujirebio Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges recalled | Fujirebio Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Lumipulse ¿-Amyloid Plasma Controls recalled by Fujirebio Diagnostics | Fujirebio Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio recalled | Fujirebio Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Lumipulse G pTau 217 Plasma Calibrators recalled | Fujirebio Diagnostics | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |