CLASS II Device recall · Reported 9 Mar 2022 · FDA Enforcement Report
Lyfo-disk, Quality control kit for culture media, sold as recalled
Microbiologics is recalling Lyfo-disk, Quality control kit for culture media, sold as: Lyfo-disk Enterococcus, distributed nationwide in the US. The recall was initiated on 6 Oct 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0713-2022
- FDA event ID
- 89451
- Recalling firm
- Microbiologics, Saint Cloud, MN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Oct 2014
- FDA classified
- 3 Mar 2022
- Reported
- 9 Mar 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 6 Oct 2014 | Recall initiated by company | |
| 3 Mar 2022 | Classified by FDA | +2705 days |
| 9 Mar 2022 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Product as listed by the FDA
LYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus faecalis derived from ATCC 51299
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
Questions about this recall
Is Lyfo-disk, Quality control kit for culture media, sold as: Lyfo-disk Enterococcus recalled?
Yes. Microbiologics recalled Lyfo-disk, Quality control kit for culture media, sold as: Lyfo-disk Enterococcus on 6 Oct 2014. The recall is Class II and its status is Terminated. Recall number Z-0713-2022.
Why was it recalled?
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Which lots are affected?
Catalog Number 0959L, UDI: 10845357023237, Lot Number (Expiration Date): 959-74-5 (10/08/2014), 959-74-8 (10/08/2014), 959-75-1 (04/22/2015), 959-75-5 (04/22/2015), 959-75-11 (04/22/2015), and 959-77-2 (10/21/2015)
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Microbiologics
All 66| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lyfo Disk, Campylobacter jejuni subsp recalled by Microbiologics | Microbiologics | Device Malfunction | Nationwide |
| Device | CLASS II | Kwik-stik, Campylobacter jejuni subsp recalled by Microbiologics | Microbiologics | Device Malfunction | Nationwide |
| Device | CLASS II | Kwik-stik, Campylobacter jejuni subsp recalled by Microbiologics | Microbiologics | Device Malfunction | Nationwide |
| Device | CLASS II | Kwik-stik, Quality control kit for culture media recalled | Microbiologics | Packaging Defects | Nationwide |
| Device | CLASS II | Lyfo-disk, M recalled by Microbiologics | Microbiologics | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |