CLASS II Device recall · Reported 29 Jul 2015 · FDA Enforcement Report

Magnus Hybrid OR Table Column (1180.01B2) used recalled

Maquet Cardiovascular Us Sales is recalling Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego, distributed nationwide in the US. The recall was initiated on 10 Jun 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSN 00048
Quantity recalled12 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2097-2015
FDA event ID
71446
Recalling firm
Maquet Cardiovascular Us Sales, Wayne, NJ
Classification
Class II
Status
Terminated
Recall initiated
10 Jun 2015
FDA classified
17 Jul 2015
Reported
29 Jul 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

10 Jun 2015Recall initiated by company
17 Jul 2015Classified by FDA+37 days
29 Jul 2015Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.

Product as listed by the FDA

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Where it was distributed

Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.

Nationwide

Questions about this recall

Why was it recalled?

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.

Which lots are affected?

SN 00048

Where was it sold?

Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.