CLASS II ONGOING Device recall · Reported 10 Dec 2025 · FDA Enforcement Report

Mammomat Inspiration recalled by Siemens Medical Solutions USA

Siemens Medical Solutions USA is recalling Mammomat Inspiration, distributed nationwide in the US. The recall was initiated on 24 Oct 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 6734, 6694, 10121, 10309, 10336, 10342;
Quantity recalled6 units (3 US, 3 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0878-2026
FDA event ID
97920
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Ongoing
Recall initiated
24 Oct 2025
FDA classified
3 Dec 2025
Reported
10 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

24 Oct 2025Recall initiated by company
3 Dec 2025Classified by FDA+40 days
10 Dec 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Product as listed by the FDA

MAMMOMAT Inspiration;

Where it was distributed

Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.

Nationwide ArkansasCaliforniaNew York

Questions about this recall

Why was it recalled?

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Which lots are affected?

Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 6734, 6694, 10121, 10309, 10336, 10342;

Where was it sold?

Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.