CLASS II ONGOING Device recall · Reported 1 May 2024 · FDA Enforcement Report

Mariner RDX Polyaxial Head recalled by Seaspine Orthopedics

Seaspine Orthopedics is recalling Mariner RDX Polyaxial Head, distributed nationwide in the US. The recall was initiated on 15 Feb 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005
Quantity recalled1579 Screw Heads

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1605-2024
FDA event ID
94164
Recalling firm
Seaspine Orthopedics, Carlsbad, CA
Classification
Class II
Status
Ongoing
Recall initiated
15 Feb 2024
FDA classified
22 Apr 2024
Reported
1 May 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals
Device type
Spinal Implants

Timeline

15 Feb 2024Recall initiated by company
22 Apr 2024Classified by FDA+67 days
1 May 2024Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.

Where it was distributed

US Nationwide distribution in the states of MI, GA, VA, IN, FL.

Nationwide FloridaGeorgiaIndianaMichiganVirginia

Questions about this recall

Is Mariner RDX Polyaxial Head recalled?

Yes. Seaspine Orthopedics recalled Mariner RDX Polyaxial Head on 15 Feb 2024. The recall is Class II and its status is Ongoing. Recall number Z-1605-2024.

Why was it recalled?

Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Which lots are affected?

REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005

Where was it sold?

US Nationwide distribution in the states of MI, GA, VA, IN, FL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Seaspine Orthopedics

All 10

Other Spinal Implants recalls

All 1,157