CLASS II ONGOING Device recall · Reported 23 Apr 2025 · FDA Enforcement Report

Medflex, REF: 1119-00; Medbank Main, REF: 1137-00; Medflex 2.0, REF recalled

CareFusion 303 is recalling Medflex, REF: 1119-00; Medbank Main, REF: 1137-00; Medflex 2.0, REF, distributed nationwide in the US. The recall was initiated on 5 Oct 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesBD Pyxis MedBank Receiving Cabinet Software Software versions: 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, 2.4.2.26 REF/UDI-DI: 1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/10885403512544; 169-104/10885403512544; 169-105/10885403512544; 169-106/10885403512544; 169-107/10885403512544; 169-108/10885403512544; 169-109/10885403512544; 169-110/10885403512544; 169-111/10885403512544; 169-112/10885403512544; 169-113/10885403512544; 169-132/10885403512568; 169-137/10885403512544; 169-138/10885403512568; 169-139/10885403512568; 169-153/10885403512568; 169-155/10885403512568; 169-156/10885403512568; 169-159/10885403512544; 169-160/10885403512544; 169-161/10885403512544; 169-162/10885403512544; 169-23/10885403512568; 169-24/10885403512568; 169-25/10885403512568; 169-26/10885403512568; 169-27/10885403512568; 169-28/10885403512568; 169-29/10885403512568; 169-30/10885403512568; 169-31/10885403512568; 169-32/10885403512568; 169-33/10885403512568; 169-34/10885403512568; 169-35/10885403512568; 169-36/10885403512568; 169-37/10885403512568; 169-38/10885403512568; 169-39/10885403512568; 169-40/10885403512568; 169-41/10885403512568; 169-42/10885403512568; 169-43/10885403512568; 169-44/10885403512568; 169-45/10885403512568; 169-46/10885403512568; 169-47/10885403512568; 169-48/10885403512568; 169-49/10885403512568; 169-50/10885403512568; 169-51/10885403512568; 169-52/10885403512568; 169-53/10885403512568; 169-54/10885403512568; 169-55/10885403512568; 169-56/10885403512568; 169-57/10885403512568; 169-58/10885403512568; 169-59/10885403512568; 169-60/10885403512568; 169-61/10885403512568; 169-62/10885403512568; 169-63/10885403512568; 169-64/10885403512568; 169-65/10885403512568; 169-66/10885403512568; 169-67/10885403512568; 169-68/10885403512568; 169-69/10885403512568; 169-70/10885403512568; 169-71/10885403512568; 169-72/10885403512568; 169-73/10885403512568; 169-74/10885403512568; 169-75/10885403512568; 169-76/10885403512568; 169-77/10885403512568; 169-78/10885403512568; 169-79/10885403512568; 169-80/10885403512568; 169-81/10885403512568; 139089-01
Quantity recalled2679

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1552-2025
FDA event ID
96029
Recalling firm
CareFusion 303, San Diego, CA
Classification
Class II
Status
Ongoing
Recall initiated
5 Oct 2023
FDA classified
11 Apr 2025
Reported
23 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

5 Oct 2023Recall initiated by company
11 Apr 2025Classified by FDA+554 days
23 Apr 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-101; BD PYXIS MEDBANK MINI 1HH-1HM- 1FM, REF: 169-102; BD PYXIS MEDBANK MINI 1HH-3HM, REF: 169-103; BD PYXIS MEDBANK MINI 2FH, REF: 169-104; BD PYXIS MEDBANK MINI 2FM, REF: 169-105; BD PYXIS MEDBANK MINI 2HH-1FH, REF: 169-106; BD PYXIS MEDBANK MINI 2HH-1FM , REF: 169-107; BD PYXIS MEDBANK MINI 2HH-2HM, REF: 169-108; BD PYXIS MEDBANK MINI 2HM-1FM, REF: 169-109; BD PYXIS MEDBANK MINI 3HH-1FG, REF: 169-110; BD PYXIS MEDBANK MINI 3HH-1HM, REF: 169-111; BD PYXIS MEDBANK MINI 4HH, REF: 169-112; BD PYXIS MEDBANK MINI 4HM, REF: 169-113; BD PYXIS MDBANK TWR MN 7HH-1HM-3FM RxNow, REF: 169-132; BD PYXIS MEDBANK MINI 1FH-1FM, REF: 169-137; BD PYXIS MEDBANK TWR MN 1FH- 10HM-1FM, REF: 169-138; BD PYXIS MEDBANK TWR MN 2HH- 4FH-2FM, REF: 169-139; OBS BD PYXIS MB TOWER MAIN, REF: 169-153; OBS BD PYXIS MB TWR RXNOW MAIN W/MED PRI, REF: 169-155; OBS BD PYXIS MB TWR RXNOW MAIN W/PAT PRI, REF: 169-156; OBS BD PYXIS MB MINI MAIN W/CUBIE DRWS, REF: 169-159; OBS BD PYXIS MB MINI MAIN W/MATRIX DRWS, REF: 169-160; OBS BD PYXIS MB MINI MAIN W/HH FALSE DRW, REF: 169-161; OBS BD PYXIS MB MINI MAIN W/FH FALSE DRW, REF: 169-162; BD PYXIS MEDBANK TWR MN 10HH- 1FH-1FM, REF: 169-23; BD PYXIS MEDBANK TWR MN 10HH- 2FH, REF: 169-24; BD PYXIS MEDBANK TWR MN 10HH- 2FM, REF: 169-25; BD PYXIS MEDBANK TWR MN 12HH- 1FF, REF: 169-26; BD PYXIS MEDBANK TWR MN 12HH- 1FH, REF: 169-27; BD PYXIS MEDBANK TWR MN 12HH- 1FM, REF: 169-28; BD PYXIS MEDBANK TWR MN 14HH, REF: 169-29; BD PYXIS MEDBANK TWR MN 1HH- 1HM-2FM-4FF, REF: 169-30; BD PYXIS MEDBANK TWR MN 2FH- 2HM-4FM, REF: 169-31; BD PYXIS MEDBANK TWR MN 2HH- 10HM-1FM, REF: 169-32; BD PYXIS MEDBANK TWR MN 2HH- 1FH-5FM, REF: 169-33; BD PYXIS MEDBANK TWR MN 2HH- 2FH-4FM, REF: 169-34; BD PYXIS MEDBANK TWR MN 2HH- 2FM-4FF, REF: 169-35; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FF, REF: 169-36; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FM, REF: 169-37; BD PYXIS MEDBANK TWR MN 2HH- 6FM, REF: 169-38; BD PYXIS MEDBANK TWR MN 2HH- 6HM-3FM, REF: 169-39; BD PYXIS MEDBANK TWR MN 2HH- 8HM-2FM, REF: 169-40; BD PYXIS MEDBANK TWR MN 2HM- 6FM, REF: 169-41; BD PYXIS MEDBANK TWR MN 3FH-4FM , REF: 169-42; BD PYXIS MEDBANK TWR MN 3HH- 7HM-2FM, REF: 169-43; BD PYXIS MEDBANK TWR MN 4HH- 10HM, REF: 169-44; BD PYXIS MEDBANK TWR MN 4HH- 1FH-2HM-3FM, REF: 169-45; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4FM, REF: 169-46; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4HM-2FM, REF: 169-47; BD PYXIS MEDBANK TWR MN 4HH- 1FH-6HM-1FM, REF: 169-48; BD PYXIS MEDBANK TWR MN 4HH- 2FH-2HM-2FM, REF: 169-49; BD PYXIS MEDBANK TWR MN 4HH- 2FH-3FM, REF: 169-50; BD PYXIS MEDBANK TWR MN 4HH- 2HM-4FM, REF: 169-51; BD PYXIS MEDBANK TWR MN 4HH- 3FH-2FM, REF: 169-52; BD PYXIS MEDBANK TWR MN 4HH- 4FH-1FM, REF: 169-53; BD PYXIS MEDBANK TWR MN 4HH- 4HM-3FM, REF: 169-54; BD PYXIS MEDBANK TWR MN 4HH-5FH, REF: 169-55; BD PYXIS MEDBANK TWR MN 4HH- 5FM, REF: 169-56; BD PYXIS MEDBANK TWR MN 4HH- 6HM-2FM, REF: 169-57; BD PYXIS MEDBANK TWR MN 4HH- 8HM-1FM, REF: 169-58; BD PYXIS MEDBANK TWR MN 4HM- 5FM, REF: 169-59; BD PYXIS MEDBANK TWR MN 5HH- 3HM-3FM, REF: 169-60; BD PYXIS MEDBANK TWR MN 5HH- 9HM, REF: 169-61; BD PYXIS MEDBANK TWR MN 6FH-1FM, REF: 169-62; BD PYXIS MEDBANK TWR MN 6HH- 1FH-2HM-2FM, REF: 169-63; BD PYXIS MEDBANK TWR MN 6HH- 1FH-3FM, REF: 169-64; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2FM, REF: 169-65; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2HM-1FM, REF: 169-66; BD PYXIS MEDBANK TWR MN 6HH-4FH, REF: 169-67; BD PYXIS MEDBANK TWR MN 6HH- 4FM, REF: 169-68; BD PYXIS MEDBANK TWR MN 6HH- 4HM-2FM, REF: 169-69; BD PYXIS MEDBANK TWR MN 6HH- 6HM-1FM, REF: 169-70; BD PYXIS MEDBANK TWR MN 6HH- 8HM, REF: 169-71; BD PYXIS MEDBANK TWR MN 7FH, REF: 169-72; BD PYXIS M

Where it was distributed

US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.

Nationwide AlabamaArizonaCaliforniaFloridaIllinoisIndianaKansasKentuckyMassachusettsMarylandNew JerseyNew YorkOhioPennsylvaniaTexasUtahWashington

Questions about this recall

Why was it recalled?

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Which lots are affected?

BD Pyxis MedBank Receiving Cabinet Software Software versions: 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, 2.4.2.26 REF/UDI-DI: 1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/10885403512544; 169-104/10885403512544; 169-105/10885403512544; 169-106/10885403512544; 169-107/10885403512544; 169-108/10885403512544; 169-109/10885403512544; 169-110/10885403512544; 169-111/10885403512544; 169-112/10885403512544; 169-113/10885403512544; 169-132/10885403512568; 169-137/10885403512544; 169-138/10885403512568; 169-139/10885403512568; 169-153/10885403512568; 169-155/10885403512568; 169-156/10885403512568; 169-159/10885403512544; 169-160/10885403512544; 169-161/10885403512544; 169-162/10885403512544; 169-23/10885403512568; 169-24/10885403512568; 169-25/10885403512568; 169-26/10885403512568; 169-27/10885403512568; 169-28/10885403512568; 169-29/10885403512568; 169-30/10885403512568; 169-31/10885403512568; 169-32/10885403512568; 169-33/10885403512568; 169-34/10885403512568; 169-35/10885403512568; 169-36/10885403512568; 169-37/10885403512568; 169-38/10885403512568; 169-39/10885403512568; 169-40/10885403512568; 169-41/10885403512568; 169-42/10885403512568; 169-43/10885403512568; 169-44/10885403512568; 169-45/10885403512568; 169-46/10885403512568; 169-47/10885403512568; 169-48/10885403512568; 169-49/10885403512568; 169-50/10885403512568; 169-51/10885403512568; 169-52/10885403512568; 169-53/10885403512568; 169-54/10885403512568; 169-55/10885403512568; 169-56/10885403512568; 169-57/10885403512568; 169-58/10885403512568; 169-59/10885403512568; 169-60/10885403512568; 169-61/10885403512568; 169-62/10885403512568; 169-63/10885403512568; 169-64/10885403512568; 169-65/10885403512568; 169-66/10885403512568; 169-67/10885403512568; 169-68/10885403512568; 169-69/10885403512568; 169-70/10885403512568; 169-71/10885403512568; 169-72/10885403512568; 169-73/10885403512568; 169-74/10885403512568; 169-75/10885403512568; 169-76/10885403512568; 169-77/10885403512568; 169-78/10885403512568; 169-79/10885403512568; 169-80/10885403512568; 169-81/10885403512568; 139089-01

Where was it sold?

US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from CareFusion 303

All 143

Other Medical Software recalls

All 932