CLASS I ONGOING Device recall · Reported 28 Jan 2026 · FDA Enforcement Report

Medline beds labeled as: Full Electric Lightweight Homecare Bed recalled

Medline Industries is recalling Medline beds labeled as: Full Electric Lightweight Homecare Bed, distributed in California, Pennsylvania. The recall was initiated on 26 Nov 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI/DI 40080196294344, All lots
Quantity recalled35694 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1015-2026
FDA event ID
98185
Recalling firm
Medline Industries, Northfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
26 Nov 2025
FDA classified
16 Jan 2026
Reported
28 Jan 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Nov 2025Recall initiated by company
16 Jan 2026Classified by FDA+51 days
28 Jan 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Product as listed by the FDA

MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L

Where it was distributed

US, CA, PA, GUAM, VI

CaliforniaPennsylvania

Questions about this recall

Why was it recalled?

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Which lots are affected?

UDI/DI 40080196294344, All lots

Where was it sold?

US, CA, PA, GUAM, VI

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.