CLASS II ONGOING Device recall · Reported 8 Apr 2026 · FDA Enforcement Report
Medline convenience kits containing Olympus biopsy valves recalled
Medline Industries is recalling Medline convenience kits containing Olympus biopsy valves: Bronch Pack, DYKE2096 Non MB, distributed nationwide in the US. The recall was initiated on 16 Feb 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1706-2026
- FDA event ID
- 98530
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 Feb 2026
- FDA classified
- 1 Apr 2026
- Reported
- 8 Apr 2026
- Type
- Voluntary: Firm initiated
- Reason category
- Foreign Material
- Device type
- Other Medical Devices
Timeline
| 16 Feb 2026 | Recall initiated by company | |
| 1 Apr 2026 | Classified by FDA | +44 days |
| 8 Apr 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
Where it was distributed
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Nationwide CaliforniaFloridaIllinoisIndianaKentuckyMassachusettsMarylandMontanaNew YorkOhioPennsylvaniaTennesseeTexas
Questions about this recall
Why was it recalled?
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Which lots are affected?
1. UDI-DI 10198459629334 (EA) 40198459629335 (CS) LOTS 26AMG895 2. UDI-DI 10198459545283 (EA) 40198459545284 (CS) LOTS 25LBJ366 26ABR365 3. UDI-DI 10195327471859 (EA) 40195327471850 (CS) LOTS 24EBA317 24EBV431 24IBD694 24JBQ783
Where was it sold?
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
6 other products · FDA event 98530This product is part of a recall event; the main page for the event is Thoracic Robots, Dynj908777b recalled by Medline Industries.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Thoracic Robots, Dynj908777b recalled by Medline Industries | Medline Industries | Foreign Material | Nationwide | |
| CLASS II | Airway Exam Kit, DYKE1796 recalled by Medline Industries | Medline Industries | Foreign Material | Nationwide | |
| CLASS II | Medline kits containing Olympus biopsy valves: Bronchoscopy recalled | Medline Industries | Foreign Material | Nationwide | |
| CLASS II | Medline kits containing Olympus biopsy valves recalled | Medline Industries | Foreign Material | Nationwide | |
| CLASS II | Medline kits containing Olympus biopsy valves: BAL KIT recalled | Medline Industries | Foreign Material | Nationwide | |
| CLASS II | Medline Anterior Hip Pack Dynj64672b recalled by Medline Industries | Medline Industries | Foreign Material | Nationwide |
More recalls from Medline Industries
All 709| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rollator, Bariatric, Basic Steel, Burgundy recalled | Medline Industries | Foreign Material | Nationwide |
| Device | CLASS II | Rollator, Bariatric, Basic Steel, Blue recalled by Medline Industries | Medline Industries | Foreign Material | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |