CLASS II ONGOING Device recall · Reported 8 Apr 2026 · FDA Enforcement Report

Thoracic Robots, Dynj908777b recalled by Medline Industries

Medline Industries is recalling Thoracic Robots, Dynj908777b, distributed nationwide in the US. The recall was initiated on 16 Feb 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735
Quantity recalled14,379 kits total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1707-2026
FDA event ID
98530
Recalling firm
Medline Industries, Northfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
16 Feb 2026
FDA classified
1 Apr 2026
Reported
8 Apr 2026
Type
Voluntary: Firm initiated
Reason category
Foreign Material

Timeline

16 Feb 2026Recall initiated by company
1 Apr 2026Classified by FDA+44 days
8 Apr 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

THORACIC ROBOTS, DYNJ908777B

Where it was distributed

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

Nationwide CaliforniaFloridaIllinoisIndianaKentuckyMassachusettsMarylandMontanaNew YorkOhioPennsylvaniaTennesseeTexas

Questions about this recall

Why was it recalled?

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Which lots are affected?

UDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735

Where was it sold?

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

6 other products · FDA event 98530

More recalls from Medline Industries