CLASS II ONGOING Device recall · Reported 24 Jun 2026 · FDA Enforcement Report
Medline Medical Procedure Kits labeled as recalled by Medline Industries
Medline Industries is recalling Medline Medical Procedure Kits labeled as: Carotid Neuro Pack, Medline Kit Number, distributed nationwide in the US. The recall was initiated on 4 May 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2500-2026
- FDA event ID
- 98994
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 4 May 2026
- FDA classified
- 17 Jun 2026
- Reported
- 24 Jun 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 4 May 2026 | Recall initiated by company | |
| 17 Jun 2026 | Classified by FDA | +44 days |
| 24 Jun 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Product as listed by the FDA
MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.
Where it was distributed
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Questions about this recall
Why was it recalled?
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Which lots are affected?
Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24EBH848; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24DBK995; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23JBP023; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23HBR067; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23FBM330; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23EBK061; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22JBG042; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22GBH864; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22DBT571; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25DMJ438; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25CMH178; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24IME720; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24HMG912; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26CMC503; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26AMB047; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 25JME109.
Where was it sold?
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
24 other products · FDA event 98994This product is part of a recall event; the main page for the event is Medline Medical Procedure Kits labeled as: Basic Heart Pack recalled.
More recalls from Medline Industries
All 709| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rollator, Bariatric, Basic Steel, Burgundy recalled | Medline Industries | Foreign Material | Nationwide |
| Device | CLASS II | Rollator, Bariatric, Basic Steel, Blue recalled by Medline Industries | Medline Industries | Foreign Material | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |