Medtronic SmartLink Software loaded on SmartSync Patient Connector recalled

Medtronic is recalling Medtronic SmartLink. The recall was initiated on 17 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
Quantity recalled3909 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2127-2025
FDA event ID
97125
Recalling firm
Medtronic, Mounds View, MN
Classification
Class II
Status
Ongoing
Recall initiated
17 Jun 2025
FDA classified
17 Jul 2025
Reported
23 Jul 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

17 Jun 2025Recall initiated by company
17 Jul 2025Classified by FDA+30 days
23 Jul 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.

Where it was distributed

Worldwide

Questions about this recall

Is Medtronic SmartLink Software loaded on SmartSync Patient Connector recalled?

Yes. Medtronic recalled Medtronic SmartLink Software loaded on SmartSync Patient Connector on 17 Jun 2025. The recall is Class II and its status is Ongoing. Recall number Z-2127-2025.

Why was it recalled?

A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.

Which lots are affected?

1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medtronic

All 148

Other Imaging Equipment recalls

All 2,336