CLASS II Device recall · Reported 25 Jan 2017 · FDA Enforcement Report

Merge Cardio recalled by Merge Healthcare

Merge Healthcare is recalling Merge Cardio, distributed nationwide in the US. The recall was initiated on 28 Mar 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVersions: 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 9.0.9, 10.0, 10.0.1, 10.1, 10.1.1, and 10.1.2
Quantity recalled188 sites potentially have the effected versions

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1046-2017
FDA event ID
75957
Recalling firm
Merge Healthcare, Hartland, WI
Classification
Class II
Status
Terminated
Recall initiated
28 Mar 2016
FDA classified
18 Jan 2017
Reported
25 Jan 2017
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction
Device type
Medical Software

Timeline

28 Mar 2016Recall initiated by company
18 Jan 2017Classified by FDA+296 days
25 Jan 2017Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review, which could result in incorrect treatment/diagnosis.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Merge Cardio software. Product Usage: Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Is Merge Cardio recalled?

Yes. Merge Healthcare recalled Merge Cardio on 28 Mar 2016. The recall is Class II and its status is Terminated. Recall number Z-1046-2017.

Why was it recalled?

If a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review, which could result in incorrect treatment/diagnosis.

Which lots are affected?

Versions: 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 9.0.9, 10.0, 10.0.1, 10.1, 10.1.1, and 10.1.2

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Merge Healthcare

All 96

Other Medical Software recalls

All 903