CLASS II ONGOING Device recall · Reported 25 Mar 2026 · FDA Enforcement Report
MICS3 Angled Sagittal Saw Attachment recalled by Howmedica Osteonics
Howmedica Osteonics is recalling MICS3 Angled Sagittal Saw Attachment, distributed nationwide in the US. The recall was initiated on 2 Feb 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1572-2026
- FDA event ID
- 98411
- Recalling firm
- Howmedica Osteonics, Mahwah, NJ
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 2 Feb 2026
- FDA classified
- 18 Mar 2026
- Reported
- 25 Mar 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 2 Feb 2026 | Recall initiated by company | |
| 18 Mar 2026 | Classified by FDA | +44 days |
| 25 Mar 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Product as listed by the FDA
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Where it was distributed
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Questions about this recall
Why was it recalled?
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Which lots are affected?
Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
Where was it sold?
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Howmedica Osteonics
All 244| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Axle Pin Assy Large Elbow Cobalt Chrome recalled by Howmedica Osteonics | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Triathlon Hinge Insert Size 4 5612-p-411 Triathlon Hinge recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio Recessed Metal Backed Patella 3045 recalled | Howmedica Osteonics | Foreign Material | Nationwide |
| Device | CLASS II | Bearing And Axle Bushing Assy Std Elbow C-m113-1-301 Krh recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio + Mobile Bearing Pcs 78-5-0910 System One Bipolar recalled | Howmedica Osteonics | Other | Nationwide |
Other Other Medical Devices recalls
All 13,884| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS II | Palacos R pro recalled by Heraeus Medical GmbH (Dental | Heraeus Medical GmbH (Dental Division) | Packaging Defects | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065 recalled | Percussionaire | Other | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV-1C recalled by Percussionaire | Percussionaire | Other | Nationwide |