CLASS II Device recall · Reported 10 Oct 2012 · FDA Enforcement Report

00780303201 MIS 12/14 Slotted Femoral Head Provisional recalled

Zimmer is recalling 00780303201 MIS 12/14 Slotted Femoral Head Provisional -3, distributed nationwide in the US. The recall was initiated on 1 Apr 2011 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots including lot 78939000 (1/7/2002) to present
Quantity recalled231,641

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2482-2012
FDA event ID
62110
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
1 Apr 2011
FDA classified
28 Sep 2012
Reported
10 Oct 2012
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

1 Apr 2011Recall initiated by company
28 Sep 2012Classified by FDA+546 days
10 Oct 2012Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +0X32 00780303203 MIS 12/14 Slotted Femoral Head Provisional +3.5X32 00780303204 MIS 12/14 Slotted Femoral Head Provisional +7X32 00780303205 MIS 12/14 Slotted Femoral Head Provisional +10.5X32 00780303214 MIS 12/14 Slotted Femoral Head Provisional +7X32 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Where it was distributed

Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp

Which lots are affected?

All lots including lot 78939000 (1/7/2002) to present

Where was it sold?

Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

11 other products · FDA event 62110

This product is part of a recall event; the main page for the event is 00789502202 VerSys Femoral Head Provisional +0 recalled by Zimmer.

More recalls from Zimmer

All 539

Other Hip & Knee Implants recalls

All 3,097