CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report

Operating table column TS7500 Mobius, Hybrid operating table column recalled

Baxter Healthcare is recalling Operating table column TS7500 Mobius, Hybrid operating table column, distributed nationwide in the US. The recall was initiated on 29 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425
Quantity recalled77 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0351-2027
FDA event ID
100117
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
29 Sep 2026
FDA classified
3 Nov 2026
Reported
11 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

29 Sep 2026Recall initiated by company
3 Nov 2026Classified by FDA+35 days
11 Nov 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Product as listed by the FDA

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Which lots are affected?

Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 100117

This product is part of a recall event; the main page for the event is TruSystem 7500 Hybrid Plus (FC) recalled by Baxter Healthcare.

More recalls from Baxter Healthcare

All 585

Other Other Medical Devices recalls

All 13,985