CLASS II Device recall · Reported 20 Aug 2014 · FDA Enforcement Report

Moore Hip Prosthesis Long Fenestrated Stem (1- Head Diameter Length recalled

Zimmer is recalling Moore Hip Prosthesis Long Fenestrated Stem (1- Head Diameter Length, distributed nationwide in the US. The recall was initiated on 1 Jul 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart 00402100838, Lot 62435853; Part 00402100841, Lot 62435854; Part 00402100843, Lot 62396506; Part 00402100844, Lot 62424468; Part 00402100848, Lot 62435852; Part 00402100848, Lot 62467218; Part 00402100848, Lot 62511381; Part 00402100851, 37108214; Part 00402100852, Lot 62511377; Part 00402100854, Lot 62511379; Part 00402101241, Lot 62352570; Part 00402101244, Lot 62424469; Part 00402101246, Lot 62400669; Part 00402101248, Lot 62400670; Part 00402101249, Lot 62467222; Part 00402101251, Lot 62470576; Part 00402101252, Lot 62435851; Part 00402101254, Lot 62424470; Part 00402101257, Lot 62524114; & Part 00402101260, Lot 62524115.
Quantity recalled68

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2215-2014
FDA event ID
68654
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
1 Jul 2014
FDA classified
13 Aug 2014
Reported
20 Aug 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

1 Jul 2014Recall initiated by company
13 Aug 2014Classified by FDA+43 days
20 Aug 2014Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Moore Hip Prosthesis Long Fenestrated Stem 38 mm (1-1/2 Inch) Head Diameter 203 mm Length. Product Usage: The Moore Hip Prosthesis is a femoral him stem that can be used where there is roentgenogrpahic evidence of a satisfactory actabulum and sufficient bone in the femoral neck to seat the prosthesis with or without bone cement.

Where it was distributed

Worldwide Distribution - US (nationwide) in the states of IN, WI, MN, OR, and TX, and one foreign consignee in Canada.

Nationwide MinnesotaOregonTexasWisconsin

Questions about this recall

Why was it recalled?

Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot

Which lots are affected?

Part 00402100838, Lot 62435853; Part 00402100841, Lot 62435854; Part 00402100843, Lot 62396506; Part 00402100844, Lot 62424468; Part 00402100848, Lot 62435852; Part 00402100848, Lot 62467218; Part 00402100848, Lot 62511381; Part 00402100851, 37108214; Part 00402100852, Lot 62511377; Part 00402100854, Lot 62511379; Part 00402101241, Lot 62352570; Part 00402101244, Lot 62424469; Part 00402101246, Lot 62400669; Part 00402101248, Lot 62400670; Part 00402101249, Lot 62467222; Part 00402101251, Lot 62470576; Part 00402101252, Lot 62435851; Part 00402101254, Lot 62424470; Part 00402101257, Lot 62524114; & Part 00402101260, Lot 62524115.

Where was it sold?

Worldwide Distribution - US (nationwide) in the states of IN, WI, MN, OR, and TX, and one foreign consignee in Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer

All 539

Other Hip & Knee Implants recalls

All 3,097