CLASS I ONGOING Device recall · Reported 22 Jan 2025 · FDA Enforcement Report
Neo-Tee T-Piece Resuscitator recalled by Mercury Enterprises
Mercury Enterprises is recalling Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, distributed nationwide in the US. The recall was initiated on 10 Dec 2024 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0821-2025
- FDA event ID
- 95967
- Recalling firm
- Mercury Enterprises, Clearwater, FL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 10 Dec 2024
- FDA classified
- 10 Jan 2025
- Reported
- 22 Jan 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- CPAP & Ventilators
Timeline
| 10 Dec 2024 | Recall initiated by company | |
| 10 Jan 2025 | Classified by FDA | +31 days |
| 22 Jan 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Product as listed by the FDA
Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842
Where it was distributed
Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaConnecticutFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesota
Questions about this recall
Is Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805 recalled?
Yes. Mercury Enterprises recalled Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805 on 10 Dec 2024. The recall is Class I and its status is Ongoing. Recall number Z-0821-2025.
Why was it recalled?
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Which lots are affected?
Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840, 2429750840; 1050841/10641043508411, 30641043508415/2429050841; 1050842/10641043508428, 30641043508422/2426750842
Where was it sold?
Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Mercury Enterprises
All 3| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Neo-Tee T-Piece Resuscitator recalled by Mercury Enterprises | Mercury Enterprises | Other | Nationwide |
| Device | CLASS II | Mercury Medical Neo-Tee T-Piece Resuscitator and T-Piece Circuit recalled | Mercury Enterprises | Lead & Heavy Metals | Nationwide |
Other CPAP & Ventilators recalls
All 552| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo-n recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS I | Astral 100/150 Resmed Astral 100/50 is a ventilator system recalled | ResMed | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical AG | Device Malfunction | Nationwide |