CLASS II Device recall · Reported 2 Mar 2022 · FDA Enforcement Report

NobelReplace Conical Connection RP recalled by Nobel Biocare Usa

Nobel Biocare Usa is recalling NobelReplace Conical Connection RP, distributed nationwide in the US. The recall was initiated on 17 Dec 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497
Quantity recalled702 pieces

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0658-2022
FDA event ID
89465
Recalling firm
Nobel Biocare Usa, Yorba Linda, CA
Classification
Class II
Status
Terminated
Recall initiated
17 Dec 2021
FDA classified
18 Feb 2022
Reported
2 Mar 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Dec 2021Recall initiated by company
18 Feb 2022Classified by FDA+63 days
2 Mar 2022Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Product as listed by the FDA

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Where it was distributed

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoFloridaGeorgiaHawaiiIdahoIllinoisIndianaKansasKentuckyMassachusettsMaineMichiganMinnesotaMississippiNorth CarolinaNew Jersey

Questions about this recall

Why was it recalled?

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Which lots are affected?

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Where was it sold?

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Nobel Biocare Usa

All 66

Other Other Medical Devices recalls

All 13,950