CLASS II ONGOING Device recall · Reported 8 Nov 2023 · FDA Enforcement Report
Nova Stat Profile Prime Plus POC Analyzer recalled by Nova Biomedical
Nova Biomedical is recalling Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, distributed nationwide in the US. The recall was initiated on 18 Sep 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0208-2024
- FDA event ID
- 93095
- Recalling firm
- Nova Biomedical, Waltham, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Sep 2023
- FDA classified
- 1 Nov 2023
- Reported
- 8 Nov 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 18 Sep 2023 | Recall initiated by company | |
| 1 Nov 2023 | Classified by FDA | +44 days |
| 8 Nov 2023 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368
Where it was distributed
Nationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam
Questions about this recall
Is Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood recalled?
Yes. Nova Biomedical recalled Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood on 18 Sep 2023. The recall is Class II and its status is Ongoing. Recall number Z-0208-2024.
Why was it recalled?
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Which lots are affected?
UDI-DI: N/A All software versions: International Except China: W1.17.1123.0 and W3.2.1123.0 V1.50.1123.0 China: V1.50.1123.0
Where was it sold?
Nationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nova Biomedical
All 19| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | StatStrip Glucose Ketone Hospital Meter System recalled | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose Ketone (mmol/L) Hospital Meter System recalled | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Prime Plus Analyzer -Intended to measure certain gases recalled | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS II | Stat Profile Prime Analyzer as follows recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |