CLASS I ONGOING Device recall · Reported 15 Oct 2025 · FDA Enforcement Report
NOxBOXi Nitric Oxide Delivery System recalled
Noxbox is recalling NOxBOXi Nitric Oxide Delivery System, distributed in Tennessee. The recall was initiated on 9 Sep 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0029-2026
- FDA event ID
- 97516
- Recalling firm
- Noxbox, Sittingbourne
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 9 Sep 2025
- FDA classified
- 6 Oct 2025
- Reported
- 15 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 9 Sep 2025 | Recall initiated by company | |
| 6 Oct 2025 | Classified by FDA | +27 days |
| 15 Oct 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Product as listed by the FDA
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Where it was distributed
Domestic: TN;
Questions about this recall
Why was it recalled?
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Which lots are affected?
Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
Where was it sold?
Domestic: TN;
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Noxbox
All 6| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Packaging Defects | Nationwide |
| Device | CLASS I | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | TN |
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS I | NOxBOXi Nitric Oxide Delivery System, Noxbox-i, Reqnoxbox-i recalled | Noxbox | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |