CLASS II Device recall · Reported 26 Jul 2017 · FDA Enforcement Report

NX 0.8950 Imaging Processing recalled by AGFA Healthcare

AGFA Healthcare is recalling Agfa Healthcare NX 0.8950 Imaging Processing, distributed in 34 states. The recall was initiated on 21 Dec 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002025, A5410000039, 2UA62524PS, 2UA62524Q2, 2UA62524RK, A5411002031, 2UA61218G4, 5CG61266X6, 2UA61218GG, 2UA63825HQ, 2UA642306H, 2UA4161JMY, 2UA52524Y5, 2UA62524RJ, 2UA62524QV, 2UA62524QB, 2UA62524RJ, 2UA62524NS, 2UA62524PJ, 2UA62524Q9, A5411000155, A5411000151, 2UA62524PH, 2UA62524QL, 2UA62524QT, A5411000153, 2UA62524Q0, 5CG6126B9J, 5CG61266YW, 5CG61266VP, 5CG61268BG, 5CG612673N, 2UA6243GNY, 2UA62424R8, 2UA62524P4, 2UA62524RY, 2UA6243GPT, 2UA62524Q6, 2UA62524PR, A5411000152, 2UA63825HR, 2UA31000Q6, 2UA31000Q8, 2UA31000R5, 2UA31000QR, 2UA319120W, 2UA3191211, 2UA31000PS, 2UA31000QX, 2UA31000Q0, 2UA31000PX, A5411000346, A5411000232, A541000199, A5411000210, A5411000209, 2UA62524P8, 5CG61266R5, 5CG5295760, 5CG5285W67, 5C5285W7P, 5CG5285W9Y, 5CG5285W5R, 5CG528W7B, A5411001163, A5411000468, 2UA52524XM, 2UA5481BTP, A5411000145, 2UA25009HM, A5411001064, A5411000159, 2UA3380VMW, 2UA3380VMG, 2UA3461QSD, A5411001039, A5411001066, A5411001065, A5411000230, A5411000177, A5411000174, CZC6349KSZ, 2UA62524RH, 2UA62524R4, 2UA6254Q7, A5411000116, A5411002095, A5411002093, A5411002092, A5411001173, A5411001172, 2UA6243GPK, 2UA6243GPL, 2UA62524QH, 2UA4150QFZ, 2UA4150QFQ, 2UA1110ZYG, 2UA3021KYP, 2UA52524XC, 2UA52524YK, 2UA52524YQ, A5411001105, MXL24220J8, A5411002030, A5411001089, A5411002058, 2UA2181HSV, A5411000238, 2UA6021JXG, 2UA511BZ2, A5411000341, A5411000340, A5411000337, 2UA5111C0V, CZ6357GKR, CZC6357GFM, A5411002051, A5411002048, A5411002047, A5411002052, A5411002049, A5411000549, 2UA62524RW, 2UA62524QS, 2UA62524P7, 2UA62524R7, 2UA62524QM, A5410001000, A5410000132, 2UA5170ZKM, 2UA5422QN4, 2UA62524PP, 2UA62524R6, 2UA1221B32, 2UA62524QD, A5411002033, 2UA62524RX, 5CG6126743, 2UA62524PB, 2UA44616DJ, 2UA63825J3, 2UA62524P6, 2UA62524QF, 2UA62524PL, 5CG61266QW, 2UA62524PN, 2UA62524RS, 2UA62524RQ, 2UA62524QJ, 5C61266Y4, A5411002064, A54110002059, CZC6349KXK, 2UA5111BZC, 2UA642306K, 2UA63825J4, 2UA64029N9, 2UA62524PD, 2UA62524P1, 2UA62524Q8, 2UA63825J6, 2UA63825JD, 2UA62524P0, 2UA63825HS, 2UA63825HN, 2UA63825J5, 2UA63825J8, 2UA61225X9, 2UA62524PT, 2UA63825J0, 2UA62524PV, 2UA5111BYZ, 5CG61266YZ, 2UA62524NY, 2UA62524RV, 5CG61266MG, 2UA63825JC, 2UA63825HX, 2UA62524QZ, 2UA62524QR, 2UA62524PF, 5CG53032V5, 2UA52223Q2, 2UA52223Q4, 2UA51519JR, 2UA62524RP, 2UA32878M, 2UA5451JKP, 2UA62524RG, 2UA62524R5, 2UA62524RN, 2UA63825HM, A5411002022, A45411000018, A5411002057, 2UA63825HZ, A5411000180, 2UA15205X9, 2UA62524NX, 2UA31000Q9, A5411000179, 2UA5451JL2, 2UA1510H6B, 2UA5311CS3, 2UA61218GT, CZC6057ZSL, A5411002024, 2UA62524RM, 2UA62524PY, 2UA1510H60, 2UA1510H5C
Quantity recalled221 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2735-2017
FDA event ID
77632
Recalling firm
AGFA Healthcare, Greenville, SC
Classification
Class II
Status
Terminated
Recall initiated
21 Dec 2016
FDA classified
14 Jul 2017
Reported
26 Jul 2017
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction

Timeline

21 Dec 2016Recall initiated by company
14 Jul 2017Classified by FDA+205 days
26 Jul 2017Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Agfa Healthcare NX 3.0.8950 Imaging Processing Software

Where it was distributed

AL, AZ, AR, CA, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV and Canada

AlabamaArkansasArizonaCaliforniaDistrict of ColumbiaFloridaGeorgiaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriMississippiNorth Carolina

Questions about this recall

Is NX 0.8950 Imaging Processing recalled?

Yes. AGFA Healthcare recalled NX 0.8950 Imaging Processing on 21 Dec 2016. The recall is Class II and its status is Terminated. Recall number Z-2735-2017.

Why was it recalled?

A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.

Which lots are affected?

2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002025, A5410000039, 2UA62524PS, 2UA62524Q2, 2UA62524RK, A5411002031, 2UA61218G4, 5CG61266X6, 2UA61218GG, 2UA63825HQ, 2UA642306H, 2UA4161JMY, 2UA52524Y5, 2UA62524RJ, 2UA62524QV, 2UA62524QB, 2UA62524RJ, 2UA62524NS, 2UA62524PJ, 2UA62524Q9, A5411000155, A5411000151, 2UA62524PH, 2UA62524QL, 2UA62524QT, A5411000153, 2UA62524Q0, 5CG6126B9J, 5CG61266YW, 5CG61266VP, 5CG61268BG, 5CG612673N, 2UA6243GNY, 2UA62424R8, 2UA62524P4, 2UA62524RY, 2UA6243GPT, 2UA62524Q6, 2UA62524PR, A5411000152, 2UA63825HR, 2UA31000Q6, 2UA31000Q8, 2UA31000R5, 2UA31000QR, 2UA319120W, 2UA3191211, 2UA31000PS, 2UA31000QX, 2UA31000Q0, 2UA31000PX, A5411000346, A5411000232, A541000199, A5411000210, A5411000209, 2UA62524P8, 5CG61266R5, 5CG5295760, 5CG5285W67, 5C5285W7P, 5CG5285W9Y, 5CG5285W5R, 5CG528W7B, A5411001163, A5411000468, 2UA52524XM, 2UA5481BTP, A5411000145, 2UA25009HM, A5411001064, A5411000159, 2UA3380VMW, 2UA3380VMG, 2UA3461QSD, A5411001039, A5411001066, A5411001065, A5411000230, A5411000177, A5411000174, CZC6349KSZ, 2UA62524RH, 2UA62524R4, 2UA6254Q7, A5411000116, A5411002095, A5411002093, A5411002092, A5411001173, A5411001172, 2UA6243GPK, 2UA6243GPL, 2UA62524QH, 2UA4150QFZ, 2UA4150QFQ, 2UA1110ZYG, 2UA3021KYP, 2UA52524XC, 2UA52524YK, 2UA52524YQ, A5411001105, MXL24220J8, A5411002030, A5411001089, A5411002058, 2UA2181HSV, A5411000238, 2UA6021JXG, 2UA511BZ2, A5411000341, A5411000340, A5411000337, 2UA5111C0V, CZ6357GKR, CZC6357GFM, A5411002051, A5411002048, A5411002047, A5411002052, A5411002049, A5411000549, 2UA62524RW, 2UA62524QS, 2UA62524P7, 2UA62524R7, 2UA62524QM, A5410001000, A5410000132, 2UA5170ZKM, 2UA5422QN4, 2UA62524PP, 2UA62524R6, 2UA1221B32, 2UA62524QD, A5411002033, 2UA62524RX, 5CG6126743, 2UA62524PB, 2UA44616DJ, 2UA63825J3, 2UA62524P6, 2UA62524QF, 2UA62524PL, 5CG61266QW, 2UA62524PN, 2UA62524RS, 2UA62524RQ, 2UA62524QJ, 5C61266Y4, A5411002064, A54110002059, CZC6349KXK, 2UA5111BZC, 2UA642306K, 2UA63825J4, 2UA64029N9, 2UA62524PD, 2UA62524P1, 2UA62524Q8, 2UA63825J6, 2UA63825JD, 2UA62524P0, 2UA63825HS, 2UA63825HN, 2UA63825J5, 2UA63825J8, 2UA61225X9, 2UA62524PT, 2UA63825J0, 2UA62524PV, 2UA5111BYZ, 5CG61266YZ, 2UA62524NY, 2UA62524RV, 5CG61266MG, 2UA63825JC, 2UA63825HX, 2UA62524QZ, 2UA62524QR, 2UA62524PF, 5CG53032V5, 2UA52223Q2, 2UA52223Q4, 2UA51519JR, 2UA62524RP, 2UA32878M, 2UA5451JKP, 2UA62524RG, 2UA62524R5, 2UA62524RN, 2UA63825HM, A5411002022, A45411000018, A5411002057, 2UA63825HZ, A5411000180, 2UA15205X9, 2UA62524NX, 2UA31000Q9, A5411000179, 2UA5451JL2, 2UA1510H6B, 2UA5311CS3, 2UA61218GT, CZC6057ZSL, A5411002024, 2UA62524RM, 2UA62524PY, 2UA1510H60, 2UA1510H5C

Where was it sold?

AL, AZ, AR, CA, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV and Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from AGFA Healthcare

All 12

Other Imaging Equipment recalls

All 2,336