CLASS II Device recall · Reported 26 Jul 2017 · FDA Enforcement Report
NX 0.8950 Imaging Processing recalled by AGFA Healthcare
AGFA Healthcare is recalling Agfa Healthcare NX 0.8950 Imaging Processing, distributed in 34 states. The recall was initiated on 21 Dec 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2735-2017
- FDA event ID
- 77632
- Recalling firm
- AGFA Healthcare, Greenville, SC
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 21 Dec 2016
- FDA classified
- 14 Jul 2017
- Reported
- 26 Jul 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 21 Dec 2016 | Recall initiated by company | |
| 14 Jul 2017 | Classified by FDA | +205 days |
| 26 Jul 2017 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Agfa Healthcare NX 3.0.8950 Imaging Processing Software
Where it was distributed
AL, AZ, AR, CA, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV and Canada
AlabamaArkansasArizonaCaliforniaDistrict of ColumbiaFloridaGeorgiaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriMississippiNorth Carolina
Questions about this recall
Is NX 0.8950 Imaging Processing recalled?
Yes. AGFA Healthcare recalled NX 0.8950 Imaging Processing on 21 Dec 2016. The recall is Class II and its status is Terminated. Recall number Z-2735-2017.
Why was it recalled?
A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.
Which lots are affected?
2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002025, A5410000039, 2UA62524PS, 2UA62524Q2, 2UA62524RK, A5411002031, 2UA61218G4, 5CG61266X6, 2UA61218GG, 2UA63825HQ, 2UA642306H, 2UA4161JMY, 2UA52524Y5, 2UA62524RJ, 2UA62524QV, 2UA62524QB, 2UA62524RJ, 2UA62524NS, 2UA62524PJ, 2UA62524Q9, A5411000155, A5411000151, 2UA62524PH, 2UA62524QL, 2UA62524QT, A5411000153, 2UA62524Q0, 5CG6126B9J, 5CG61266YW, 5CG61266VP, 5CG61268BG, 5CG612673N, 2UA6243GNY, 2UA62424R8, 2UA62524P4, 2UA62524RY, 2UA6243GPT, 2UA62524Q6, 2UA62524PR, A5411000152, 2UA63825HR, 2UA31000Q6, 2UA31000Q8, 2UA31000R5, 2UA31000QR, 2UA319120W, 2UA3191211, 2UA31000PS, 2UA31000QX, 2UA31000Q0, 2UA31000PX, A5411000346, A5411000232, A541000199, A5411000210, A5411000209, 2UA62524P8, 5CG61266R5, 5CG5295760, 5CG5285W67, 5C5285W7P, 5CG5285W9Y, 5CG5285W5R, 5CG528W7B, A5411001163, A5411000468, 2UA52524XM, 2UA5481BTP, A5411000145, 2UA25009HM, A5411001064, A5411000159, 2UA3380VMW, 2UA3380VMG, 2UA3461QSD, A5411001039, A5411001066, A5411001065, A5411000230, A5411000177, A5411000174, CZC6349KSZ, 2UA62524RH, 2UA62524R4, 2UA6254Q7, A5411000116, A5411002095, A5411002093, A5411002092, A5411001173, A5411001172, 2UA6243GPK, 2UA6243GPL, 2UA62524QH, 2UA4150QFZ, 2UA4150QFQ, 2UA1110ZYG, 2UA3021KYP, 2UA52524XC, 2UA52524YK, 2UA52524YQ, A5411001105, MXL24220J8, A5411002030, A5411001089, A5411002058, 2UA2181HSV, A5411000238, 2UA6021JXG, 2UA511BZ2, A5411000341, A5411000340, A5411000337, 2UA5111C0V, CZ6357GKR, CZC6357GFM, A5411002051, A5411002048, A5411002047, A5411002052, A5411002049, A5411000549, 2UA62524RW, 2UA62524QS, 2UA62524P7, 2UA62524R7, 2UA62524QM, A5410001000, A5410000132, 2UA5170ZKM, 2UA5422QN4, 2UA62524PP, 2UA62524R6, 2UA1221B32, 2UA62524QD, A5411002033, 2UA62524RX, 5CG6126743, 2UA62524PB, 2UA44616DJ, 2UA63825J3, 2UA62524P6, 2UA62524QF, 2UA62524PL, 5CG61266QW, 2UA62524PN, 2UA62524RS, 2UA62524RQ, 2UA62524QJ, 5C61266Y4, A5411002064, A54110002059, CZC6349KXK, 2UA5111BZC, 2UA642306K, 2UA63825J4, 2UA64029N9, 2UA62524PD, 2UA62524P1, 2UA62524Q8, 2UA63825J6, 2UA63825JD, 2UA62524P0, 2UA63825HS, 2UA63825HN, 2UA63825J5, 2UA63825J8, 2UA61225X9, 2UA62524PT, 2UA63825J0, 2UA62524PV, 2UA5111BYZ, 5CG61266YZ, 2UA62524NY, 2UA62524RV, 5CG61266MG, 2UA63825JC, 2UA63825HX, 2UA62524QZ, 2UA62524QR, 2UA62524PF, 5CG53032V5, 2UA52223Q2, 2UA52223Q4, 2UA51519JR, 2UA62524RP, 2UA32878M, 2UA5451JKP, 2UA62524RG, 2UA62524R5, 2UA62524RN, 2UA63825HM, A5411002022, A45411000018, A5411002057, 2UA63825HZ, A5411000180, 2UA15205X9, 2UA62524NX, 2UA31000Q9, A5411000179, 2UA5451JL2, 2UA1510H6B, 2UA5311CS3, 2UA61218GT, CZC6057ZSL, A5411002024, 2UA62524RM, 2UA62524PY, 2UA1510H60, 2UA1510H5C
Where was it sold?
AL, AZ, AR, CA, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV and Canada
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from AGFA Healthcare
All 12| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Brand name: Agfa Digital Radiography X- Ray system DR recalled | AGFA Healthcare | Other | Nationwide |
| Device | CLASS II | Impax Cardiovascular The Impax CV Reporting module consists recalled | AGFA Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | DX-D600 - DXD Imaging Package recalled by AGFA Healthcare | AGFA Healthcare | Packaging Defects | 24 states |
| Device | CLASS II | AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is recalled | AGFA Healthcare | Other | Nationwide |
| Device | CLASS II | Impax, Pacs, Picture Archive and Communications System recalled | AGFA Healthcare | Device Malfunction | 13 states |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device is recalled | Canon Medical System, USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederland B.v | Lead & Heavy Metals |