CLASS II ONGOING Device recall · Reported 14 Dec 2022 · FDA Enforcement Report

Omnipod 5 Automated Insulin Delivery System Product recalled by Insulet

Insulet is recalling Omnipod 5 Automated Insulin Delivery System Product, distributed nationwide in the US. The recall was initiated on 14 Nov 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesALL serial numbers/ controller lots are currently in scope of this recall
Quantity recalled129,086

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0423-2023
FDA event ID
91127
Recalling firm
Insulet, Acton, MA
Classification
Class II
Status
Ongoing
Recall initiated
14 Nov 2022
FDA classified
7 Dec 2022
Reported
14 Dec 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Nov 2022Recall initiated by company
7 Dec 2022Classified by FDA+23 days
14 Dec 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

Where it was distributed

Domestic distribution nationwide.

Nationwide

Questions about this recall

Why was it recalled?

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Which lots are affected?

ALL serial numbers/ controller lots are currently in scope of this recall

Where was it sold?

Domestic distribution nationwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Insulet

All 21

Other Insulin Pumps & Glucose Monitors recalls

All 467