CLASS II ONGOING Device recall · Reported 14 Dec 2022 · FDA Enforcement Report
Omnipod 5 Automated Insulin Delivery System Product recalled by Insulet
Insulet is recalling Omnipod 5 Automated Insulin Delivery System Product, distributed nationwide in the US. The recall was initiated on 14 Nov 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0423-2023
- FDA event ID
- 91127
- Recalling firm
- Insulet, Acton, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Nov 2022
- FDA classified
- 7 Dec 2022
- Reported
- 14 Dec 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 14 Nov 2022 | Recall initiated by company | |
| 7 Dec 2022 | Classified by FDA | +23 days |
| 14 Dec 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM
Where it was distributed
Domestic distribution nationwide.
Questions about this recall
Why was it recalled?
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Which lots are affected?
ALL serial numbers/ controller lots are currently in scope of this recall
Where was it sold?
Domestic distribution nationwide.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Insulet
All 21| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Omnipod (Eros) Pod recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod DASH Pod recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod 5 Pods recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS I | Omnipod 5 Pods recalled by Insulet | Insulet | Packaging Defects | Nationwide |
| Device | CLASS II | The failure occurs if a user on the Omnipod 5 iOS recalled by Insulet | Insulet | Device Malfunction | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 467| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction | |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose and B- Ketone meter recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |