CLASS II ONGOING Device recall · Reported 28 May 2025 · FDA Enforcement Report

PhenoMATRIX recalled by Copan WASP

Copan Wasp is recalling PhenoMATRIX, distributed nationwide in the US. The recall was initiated on 6 May 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll software versions
Quantity recalled14

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1812-2025
FDA event ID
96765
Recalling firm
Copan Wasp, Brescia, N/A
Classification
Class II
Status
Ongoing
Recall initiated
6 May 2025
FDA classified
21 May 2025
Reported
28 May 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

6 May 2025Recall initiated by company
21 May 2025Classified by FDA+15 days
28 May 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

Product as listed by the FDA

PhenoMATRIX

Where it was distributed

US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.

Nationwide ArkansasCaliforniaIllinoisKansasMassachusettsMarylandNorth CarolinaNew JerseyNevadaNew YorkPennsylvaniaTexasUtah

Questions about this recall

Why was it recalled?

AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

Which lots are affected?

All software versions

Where was it sold?

US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Copan Wasp

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIBrand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 recalledCopan WaspDevice MalfunctionNationwide

Other Other Medical Devices recalls

All 13,908