CLASS II ONGOING Device recall · Reported 24 Jun 2026 · FDA Enforcement Report

Philips Allura Xper with the following recalled

Philips Medical Systems Nederl is recalling Philips Allura Xper with the following. The recall was initiated on 18 May 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll codes

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2412-2026
FDA event ID
99223
Recalling firm
Philips Medical Systems Nederl, Best
Classification
Class II
Status
Ongoing
Recall initiated
18 May 2026
FDA classified
16 Jun 2026
Reported
24 Jun 2026
Type
FDA Mandated
Reason category
Other Reasons

Timeline

18 May 2026Recall initiated by company
16 Jun 2026Classified by FDA+29 days
24 Jun 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

Product as listed by the FDA

Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # 722010 for Allura Xper FD10; Model # 722011 for Allura Xper FD10/10; Model # 722012 for Allura Xper FD20; Model # 722013 for Allura Xper FD20 Biplane; Model # 722015 for Allura Xper FD20 OR Table; Model # 722019 for Allura Xper FD10/10 OR Table; Model # 722020 for Allura Xper FD20 Biplane OR Table; Model # 722022 for Allura Xper FD10 OR Table; Model # 722023 for Allura Xper FD20 OR Table; Model # 722025 for Allura Xper FD20 Biplane OR Table; Model # 722026 for Allura Xper FD10; Model # 722027 for Allura Xper FD10/10; Model # 722028 for Allura Xper FD20; Model # 722029 for Allura Xper FD20/10; Model #722033 for Allura Xper FD10 OR Table; Model #722035 for Allura Xper FD20 OR Table; Model #722038 for Allura Xper FD20/20; Model #722039 for Allura Xper FD20/20 OR Table; Model #722058 for Allura Xper FD20/15; Model #722059 for Allura Xper FD20/15 OR Table.

Where it was distributed

US and ROW.

Questions about this recall

Why was it recalled?

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

Which lots are affected?

All codes

Where was it sold?

US and ROW.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99223

This product is part of a recall event; the main page for the event is Philips MultiDiagnost Eleva with the following recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIPhilips MultiDiagnost Eleva with the following recalledPhilips Medical Systems NederlOther

More recalls from Philips Medical Systems Nederl

All 200

Other Other Medical Devices recalls

All 13,884