CLASS II Device recall · Reported 30 May 2018 · FDA Enforcement Report

Philips HeartStart (HS1) Onsite/Home AED recalled

Philips Electronics North America is recalling Philips HeartStart (HS1) Onsite/Home AED, distributed nationwide in the US. The recall was initiated on 14 Mar 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesManufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges: Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375). In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.
Quantity recalledTotal for FRx and HS1 - 660,519 defibrillators

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1914-2018
FDA event ID
79343
Recalling firm
Philips Electronics North America, Bothell, WA
Classification
Class II
Status
Terminated
Recall initiated
14 Mar 2018
FDA classified
23 May 2018
Reported
30 May 2018
Type
FDA Mandated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

14 Mar 2018Recall initiated by company
23 May 2018Classified by FDA+70 days
30 May 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

Product as listed by the FDA

Philips HeartStart (HS1) Onsite/Home AED

Where it was distributed

Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.

Nationwide GeorgiaPuerto Rico

Questions about this recall

Is Philips HeartStart (HS1) Onsite/Home AED recalled?

Yes. Philips Electronics North America recalled Philips HeartStart (HS1) Onsite/Home AED on 14 Mar 2018. The recall is Class II and its status is Terminated. Recall number Z-1914-2018.

Why was it recalled?

An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

Which lots are affected?

Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges: Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375). In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.

Where was it sold?

Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.