CLASS II Device recall · Reported 15 Jun 2016 · FDA Enforcement Report

The Plum 360 is a large volume infuser capable recalled by Hospira

Hospira is recalling The Plum 360 is a large volume infuser capable of delivering fluids for a variety of. The recall was initiated on 25 Mar 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPLum 360 devices with the following list number: 300100405, 300100406, 300100605 , 300101305 , 300102705 , 300105405 and 300106507.
Quantity recalled1490 infusion pumps.

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1832-2016
FDA event ID
73799
Recalling firm
Hospira, Lake Forest, IL
Classification
Class II
Status
Terminated
Recall initiated
25 Mar 2016
FDA classified
3 Jun 2016
Reported
15 Jun 2016
Reason category
Foreign Material
Device type
Medical Software

Timeline

25 Mar 2016Recall initiated by company
3 Jun 2016Classified by FDA+70 days
15 Jun 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

The Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions. The Plum 360 infuser can deliver fluids over a broad range of infusion rates and is capable of Concurrent delivery from one or more rigid or flexible fluid containers. The Plum 360 infuser features Concurrent, Secondary, and Piggyback infusions. A positive valving cassette allows two lines to be delivered at independent rates. The volume to be infused (VTBI) is delivered through one line to a patient. The two lines can be delivered in Concurrent mode (together) or Piggyback mode (one after another) without raising or lowering I.V. bags. The Plum 360 infuser also enables fluid pathway troubleshooting such as removing proximal air in line, without is disconnecting the patient line. The Plum 360 can act as a stand-alone infuser, or in conjunction with the Hospira MedNet software to provide medication safety software at the point of care, with customized drug libraries to support hospital defined protocols by clinical care area. In such a configuration, the Plum 360 infuser can communicate with systems on the network via Ethernet or state of the art wireless communication using an 802.11 a/b/g/n/, 2.4 GHz/5 GHz dual-band radio. The Plum 360 infuser and Hospira MedNet software interface with other hospital systems such as Electronic Health records, Electronic Medication Administration Records, Bar Code Point of Care, Real Time Location Services, and there systems. Each infuser includes a Connectivity Engine (CE) which provides both wired Ethernet and wireless 802.11 a/b/g/ networking capabilities. The Plum 360 infuser interfaces with Hospira MedNet application software to download drug library and infuser software updates and enable auto-programming of the infuser.The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Where it was distributed

U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.

Questions about this recall

Why was it recalled?

Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.

Which lots are affected?

PLum 360 devices with the following list number: 300100405, 300100406, 300100605 , 300101305 , 300102705 , 300105405 and 300106507.

Where was it sold?

U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 73799

This product is part of a recall event; the main page for the event is The Plum A+ is a dual-line volumetric infusion system recalled.

More recalls from Hospira

All 225

Other Medical Software recalls

All 939