CLASS II ONGOING Device recall · Reported 6 Aug 2025 · FDA Enforcement Report

Precision Sampling Set recalled by Capnia

Capnia is recalling Precision Sampling Set, distributed nationwide in the US. The recall was initiated on 21 Apr 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesMonitor UDI-DI: 00858481006017. Lot (C20206): 231220
Quantity recalled2160

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2198-2025
FDA event ID
97184
Recalling firm
Capnia, Foster City, CA
Classification
Class II
Status
Ongoing
Recall initiated
21 Apr 2025
FDA classified
29 Jul 2025
Reported
6 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects
Device type
Patient Monitors

Timeline

21 Apr 2025Recall initiated by company
29 Jul 2025Classified by FDA+99 days
6 Aug 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Where it was distributed

US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.

Nationwide FloridaIowaMassachusettsPennsylvaniaTexasUtah

Questions about this recall

Why was it recalled?

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

Which lots are affected?

Monitor UDI-DI: 00858481006017. Lot (C20206): 231220

Where was it sold?

US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.