CLASS II ONGOING Device recall · Reported 25 Sep 2024 · FDA Enforcement Report
Prgm 2090 Programmer-portable Crm recalled by Medtronic
Medtronic is recalling Prgm 2090 Programmer-portable Crm. The recall was initiated on 11 Jul 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3152-2024
- FDA event ID
- 95176
- Recalling firm
- Medtronic, Mounds View, MN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 11 Jul 2024
- FDA classified
- 13 Sep 2024
- Reported
- 25 Sep 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 11 Jul 2024 | Recall initiated by company | |
| 13 Sep 2024 | Classified by FDA | +64 days |
| 25 Sep 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT
Where it was distributed
Worldwide distribution.
Questions about this recall
Why was it recalled?
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Which lots are affected?
2090 Programmers with serial number prefixes PKK0 and PKK1: a) GTIN: 00613994136480, 00613994136497, 00613994136503, 00613994136510, 00613994136527, 00613994136534, 00613994136565, 00613994136596, 00613994136619, 00613994136626, 00613994175076, 00613994175083, 00613994175090, 00613994175106, 00613994175120, 00613994175137, 00613994175144, 00613994175151, 00613994175168, 00613994175175, 00613994175182, 00613994175205, 00613994175212, 00613994175229, 00613994175236, 00613994175243, 00613994175250, 00613994175267, 00613994175281, 00613994175298, 00613994175304, 00613994175311, 00613994175328, 00613994175335, 00613994175342, 00613994175366, 00613994175373, 00613994175380, 00613994175397, 00613994175403, 00613994175410, 00613994175434, 00613994175465, 00613994175489, 00613994175540, 00613994175588, 00613994175595, 00613994203182, 00613994360922, 00613994434883, 00613994434890, 00613994434944, 00613994435002, 00613994435255, 00613994435316, 00613994580863, 00613994580894, 00613994580931, 00613994580962, 00613994581013, 00613994581051, 00613994581075, 00613994581105, 00613994581150, 00613994581167, 00613994581174, 00613994581181, 00613994581204, 00613994581211, 00613994581228, 00613994581235, 00613994581242, 00613994581549, 00613994581570, 00613994581587, 00613994581594, 00613994581624, 00643169440128, 00643169440166, 00643169440173, 00643169440197, 00643169440203, 00643169440937, 00643169537750, 00643169537774, 00643169537811, 00643169537828, 00643169537835, 00643169537842, 00643169537859, 00643169537866, 00643169537880, 00643169537910, 00643169537927, 00643169538016, 00643169629004, 00643169629011, 00643169629028, 00643169629035, 00643169629042, 00643169629066, 00721902246227, 00721902246234, 00721902246241, 00721902246265, 00721902246289, 00721902246302, 00721902246326, 00721902246333; b) GTIN: 00721902171222; c) GTIN: 00721902171079; d) GTIN: 00721902824692; e) GTIN: 00721902059100; f) GTIN: 00721902171154; g) GTIN: 00721902246357; h) GTIN: 00721902246364; i) GTIN: 00721902480201; j) GTIN: 00721902246159, 00721902246173; k) GTIN: 00721902246203; l) GTIN: 00721902841903; m) GTIN: 00721902246272; n) GTIN: 00613994127631, 00613994435354, 00613994581549, 00885074301440, 00885074310596, 00885074310633, 00885074310657, 00885074310930, 00885074312866, 00885074312873, 00885074313047, 00885074313108, 00885074313115, 00885074313139, 00885074313146, 00885074313153, 00885074313160; o) GTIN: 00885074313191; p) GTIN: 00885074313177; q) GTIN: 00885074313184; r) GTIN: 00885074314921; s) GTIN: 00885074313009; t) GTIN: 00885074357416; u) GTIN: 00885074313122
Where was it sold?
Worldwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Medtronic
All 149| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | OmniaSecure Model 3930M, Lead 3930M74 74 CM DF4 MR 3930M74 Lead recalled | Medtronic | Other | Nationwide |
| Device | CLASS II | Sphere-9 Catheter recalled by Medtronic | Medtronic | Other | MN |
| Device | CLASS II | Abre Venous Self-Expanding Stent System recalled by Medtronic | Medtronic | Other | Nationwide |
| Device | CLASS II | CareLink SmartSync Device Manager recalled by Medtronic | Medtronic | Device Malfunction | Nationwide |
| Device | CLASS II | CareLink SmartSync Patient Connector recalled by Medtronic | Medtronic | Device Malfunction | Nationwide |
Other Patient Monitors recalls
All 1,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Passport Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | T1 Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | cPM Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | BeneVision Series Telemetry Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | Panorama and Telemetry Monitoring System recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |