CLASS II Device recall · Reported 10 Jun 2015 · FDA Enforcement Report

PRI Femoral Impactor Block for knee prosthesis surgery recalled

Zimmer is recalling PRI Femoral Impactor Block for knee prosthesis surgery, distributed nationwide in the US. The recall was initiated on 30 Apr 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem No. 00-5901-032-00 Lots: 56572845 56574197 56574534 56574775 56575838 56573249 56574198 56574535 56575827 56575839 56573282 56574325 56574536 56575828 56575840 56573283 56574326 56574604 56575829 56576089 56573284 56574327 56574605 56575830 56576090 56573285 56574328 56574606 56575831 56576091 56574130 56574329 56574607 56575832 56576421 56574135 56574530 56574608 56575833 56576429 56574136 56574531 56574617 56575836 56576431 56574137 56574532 56574714 56575837 56576544 56574196 56574533 56574774
Quantity recalled1,689

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1696-2015
FDA event ID
71250
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
30 Apr 2015
FDA classified
29 May 2015
Reported
10 Jun 2015
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Foreign Material

Timeline

30 Apr 2015Recall initiated by company
29 May 2015Classified by FDA+29 days
10 Jun 2015Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument

Where it was distributed

Worldwide Distribution-US (nationwide) including PR and the states of AL, AZ, CA, CO, FL, IL, IN, KS, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, and the countries of Canada, Korea, Singapore, Australia, Japan, Malaysia, India, Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Iceland, Iran, Israel, Italy, Norway, Russian Federation, Saudi Arabia, South America, Spain, Sweden, Switzerland, United Kingdom, and United Arab Emirates.

Nationwide AlabamaArizonaCaliforniaColoradoFloridaIllinoisIndianaKansasMassachusettsMichiganMinnesotaMissouriNorth CarolinaNew JerseyNew YorkOhioOklahomaPennsylvaniaPuerto RicoTennessee

Questions about this recall

Why was it recalled?

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

Which lots are affected?

Item No. 00-5901-032-00 Lots: 56572845 56574197 56574534 56574775 56575838 56573249 56574198 56574535 56575827 56575839 56573282 56574325 56574536 56575828 56575840 56573283 56574326 56574604 56575829 56576089 56573284 56574327 56574605 56575830 56576090 56573285 56574328 56574606 56575831 56576091 56574130 56574329 56574607 56575832 56576421 56574135 56574530 56574608 56575833 56576429 56574136 56574531 56574617 56575836 56576431 56574137 56574532 56574714 56575837 56576544 56574196 56574533 56574774

Where was it sold?

Worldwide Distribution-US (nationwide) including PR and the states of AL, AZ, CA, CO, FL, IL, IN, KS, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, and the countries of Canada, Korea, Singapore, Australia, Japan, Malaysia, India, Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Iceland, Iran, Israel, Italy, Norway, Russian Federation, Saudi Arabia, South America, Spain, Sweden, Switzerland, United Kingdom, and United Arab Emirates.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.