CLASS II ONGOING Device recall · Reported 10 Jul 2024 · FDA Enforcement Report

Prucka 3 Amplifier recalled by GE Medical Systems

GE Medical Systems Information Technologies is recalling Prucka 3 Amplifier, distributed nationwide in the US. The recall was initiated on 22 Apr 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI N/A for the Prucka 3 Amplifier-not a medical device. CardioLab and ComboLab Systems GTIN #00195278507044 and #00195278507051. Prucka 3 Amplifier serial numbers: SVS23400020SA, SVS23400024SA, SVS23360017SA, SVS23400003SA, SVS23470003SA, SVS23350005SA, SVS23330008SA, SVS23350001SA, SVS23350003SA, SVS23360018SA, SVS23520022SA, SVS23350008SA, SVS23280009SA, SVS23350021SA, SVS23470016SA, SVS23470010SA, SVS23360013SA, SVS23280002SA, SVS23400026SA, SVS23400039SA, SVS23400021SA, SVS23330006SA, SVS23360014SA, SVS23360016SA, SVS23460020SA, SVS23470022SA, SVS23480007SA, SVS23360010SA, SVS23350024SA, SVS24020002SA, SVS24020004SA, SVS24020005SA, SVS24020006SA, SVS24020007SA, SVS24020011SA, SVS23460002SA, SVS23350004SA, SVS23350006SA, SVS23350007SA, SVS23480003SA, SVS23400008SA, SVS23210005SA, SVS23360007SA, SVS23360002SA, SVS23360012SA, SVS23470004SA, SVS23520018SA, SVS23520020SA, SVS24020016SA, SVS23460001SA, SVS23470006SA, SVS23470008SA, SVS23350002SA, SVS23350009SA, SVS23350025SA, SVS23360003SA, SVS23360005SA, SVS23360006SA, SVS23360011SA, SVS23400023SA, SVS23400030SA, SVS23400031SA, SVS23400032SA, SVS23400037SA, SVS23400038SA, SVS23460007SA, SVS23400015SA, SVS23350011SA, SVS23350017SA, SVS23470017SA, SVS23470018SA, SVS23470020SA, SVS23460013SA, SVS23460014SA, SVS23480002SA, SVS23480005SA, SVS23480010SA, SVS23480011SA, SVS23480013SA, SVS23480015SA, SVS23520012SA, SVS23280008SA, SVS23290004SA, SVS23330004SA, SVS23350014SA, SVS23350022SA, SVS23350023SA, SVS23460006SA, SVS23400022SA, SVS23350012SA, SVS23350016SA, SVS23480004SA, SVS23470007SA, SVS23400025SA, SVS23460003SA, SVS23460004SA, SVS23350015SA, SVS23320002SA, SVS23470013SA, SVS23350020SA, SVS23330005SA, SVS23470009SA, SVS23470011SA, SVS23470019SA, SVS23470012SA, SVS23460016SA, SVS23290001SA, SVS23350013SA, SVS23460017SA, SVS23480012SA, SVS23480016SA, SVS23330007SA, SVS23360008SA, SVS23360020SA, SVS23400005SA, SVS23400009SA, SVS23400010SA, SVS23400019SA, SVS23460008SA, SVS23280005SA, SVS23400004SA, SVS23460005SA, SVS23400034SA, SVS23400035SA, SVS23400036SA, SVS23400033SA, SVS23400018SA, SVS23470005SA, SVS23400028SA, SVS23400029SA, SVS23360019SA, SVS23400002SA, SVS23360004SA, SVS23400007SA, SVS23400011SA, SVS23400006SA, SVS23350018SA, SVS23330003SA, SVS23290003SA, SVS23290002SA, SVS23330009SA, SVS23350019SA, SVS23360015SA, SVS23470001SA, and SVS23470002SA.
Quantity recalled145 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2258-2024
FDA event ID
94717
Recalling firm
GE Medical Systems Information Technologies, Milwaukee, WI
Classification
Class II
Status
Ongoing
Recall initiated
22 Apr 2024
FDA classified
3 Jul 2024
Reported
10 Jul 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

22 Apr 2024Recall initiated by company
3 Jul 2024Classified by FDA+72 days
10 Jul 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

Product as listed by the FDA

Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).

Where it was distributed

Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.

Nationwide AlaskaArizonaCaliforniaColoradoFloridaGeorgiaIllinoisKentuckyMassachusettsMaineMinnesotaMissouriNorth CarolinaNew JerseyNew YorkOhioOregonPennsylvaniaTennesseeTexas

Questions about this recall

Why was it recalled?

Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

Which lots are affected?

UDI N/A for the Prucka 3 Amplifier-not a medical device. CardioLab and ComboLab Systems GTIN #00195278507044 and #00195278507051. Prucka 3 Amplifier serial numbers: SVS23400020SA, SVS23400024SA, SVS23360017SA, SVS23400003SA, SVS23470003SA, SVS23350005SA, SVS23330008SA, SVS23350001SA, SVS23350003SA, SVS23360018SA, SVS23520022SA, SVS23350008SA, SVS23280009SA, SVS23350021SA, SVS23470016SA, SVS23470010SA, SVS23360013SA, SVS23280002SA, SVS23400026SA, SVS23400039SA, SVS23400021SA, SVS23330006SA, SVS23360014SA, SVS23360016SA, SVS23460020SA, SVS23470022SA, SVS23480007SA, SVS23360010SA, SVS23350024SA, SVS24020002SA, SVS24020004SA, SVS24020005SA, SVS24020006SA, SVS24020007SA, SVS24020011SA, SVS23460002SA, SVS23350004SA, SVS23350006SA, SVS23350007SA, SVS23480003SA, SVS23400008SA, SVS23210005SA, SVS23360007SA, SVS23360002SA, SVS23360012SA, SVS23470004SA, SVS23520018SA, SVS23520020SA, SVS24020016SA, SVS23460001SA, SVS23470006SA, SVS23470008SA, SVS23350002SA, SVS23350009SA, SVS23350025SA, SVS23360003SA, SVS23360005SA, SVS23360006SA, SVS23360011SA, SVS23400023SA, SVS23400030SA, SVS23400031SA, SVS23400032SA, SVS23400037SA, SVS23400038SA, SVS23460007SA, SVS23400015SA, SVS23350011SA, SVS23350017SA, SVS23470017SA, SVS23470018SA, SVS23470020SA, SVS23460013SA, SVS23460014SA, SVS23480002SA, SVS23480005SA, SVS23480010SA, SVS23480011SA, SVS23480013SA, SVS23480015SA, SVS23520012SA, SVS23280008SA, SVS23290004SA, SVS23330004SA, SVS23350014SA, SVS23350022SA, SVS23350023SA, SVS23460006SA, SVS23400022SA, SVS23350012SA, SVS23350016SA, SVS23480004SA, SVS23470007SA, SVS23400025SA, SVS23460003SA, SVS23460004SA, SVS23350015SA, SVS23320002SA, SVS23470013SA, SVS23350020SA, SVS23330005SA, SVS23470009SA, SVS23470011SA, SVS23470019SA, SVS23470012SA, SVS23460016SA, SVS23290001SA, SVS23350013SA, SVS23460017SA, SVS23480012SA, SVS23480016SA, SVS23330007SA, SVS23360008SA, SVS23360020SA, SVS23400005SA, SVS23400009SA, SVS23400010SA, SVS23400019SA, SVS23460008SA, SVS23280005SA, SVS23400004SA, SVS23460005SA, SVS23400034SA, SVS23400035SA, SVS23400036SA, SVS23400033SA, SVS23400018SA, SVS23470005SA, SVS23400028SA, SVS23400029SA, SVS23360019SA, SVS23400002SA, SVS23360004SA, SVS23400007SA, SVS23400011SA, SVS23400006SA, SVS23350018SA, SVS23330003SA, SVS23290003SA, SVS23290002SA, SVS23330009SA, SVS23350019SA, SVS23360015SA, SVS23470001SA, and SVS23470002SA.

Where was it sold?

Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.