CLASS II Device recall · Reported 29 Jan 2014 · FDA Enforcement Report
Puritan Bennett 840 Series Ventilator PB840 Operators recalled
Nellcor Puritan Bennett Inc. ( is recalling Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02, distributed nationwide in the US. The recall was initiated on 25 Jan 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0821-2014
- FDA event ID
- 67163
- Recalling firm
- Nellcor Puritan Bennett Inc. (, Boulder, CO
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 25 Jan 2013
- FDA classified
- 23 Jan 2014
- Reported
- 29 Jan 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 25 Jan 2013 | Recall initiated by company | |
| 23 Jan 2014 | Classified by FDA | +363 days |
| 29 Jan 2014 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients.
Where it was distributed
Worldwide Distribution: US (nationwide) and Internationally to: Albania, Argentina, Australia, Bahrain, Belgium, Belize, Bermuda Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, The former Yugoslav, Malaysia, Mexico, Moldova, Republic of, Montenegro, Morocco, Netherlands, Nicaragua, Northern Mariana Islands, Oman, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukrainian SSR, United Arab Emirates (UAE) United Kingdom, Uruguay, Venezuela and Yugoslavia.
Questions about this recall
Is Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 recalled?
Yes. Nellcor Puritan Bennett Inc. ( recalled Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 on 25 Jan 2013. The recall is Class II and its status is Terminated. Recall number Z-0821-2014.
Why was it recalled?
Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
Which lots are affected?
PB840 Operators Manual Addendum: PN 066009A 09/02
Where was it sold?
Worldwide Distribution: US (nationwide) and Internationally to: Albania, Argentina, Australia, Bahrain, Belgium, Belize, Bermuda Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, The former Yugoslav, Malaysia, Mexico, Moldova, Republic of, Montenegro, Morocco, Netherlands, Nicaragua, Northern Mariana Islands, Oman, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukrainian SSR, United Arab Emirates (UAE) United Kingdom, Uruguay, Venezuela and Yugoslavia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nellcor Puritan Bennett Inc. (
All 11| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Covidien Puritan Bennett 980 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Other | Nationwide |
| Device | CLASS I | Covidien Puritan Bennett 980 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Lead & Heavy Metals | 17 states |
| Device | CLASS II | Vital Sync RMS 0 when loaded with v2.4 recalled | Nellcor Puritan Bennett Inc. ( | Packaging Defects | 8 states |
| Device | CLASS II | Vital Sync" VPMP & IM recalled by Nellcor Puritan Bennett Inc. ( | Nellcor Puritan Bennett Inc. ( | Packaging Defects | 8 states |
| Device | CLASS II | Puritan Bennett 840 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Device Malfunction | Nationwide |
Other CPAP & Ventilators recalls
All 552| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo-n recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS I | Astral 100/150 Resmed Astral 100/50 is a ventilator system recalled | ResMed | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical AG | Device Malfunction | Nationwide |