CLASS II Device recall · Reported 16 Feb 2022 · FDA Enforcement Report

QC Sets and Panels, Kwik-stik Ast-gp (4 Strains) QC Set recalled

Microbiologics is recalling QC Sets and Panels, Kwik-stik Ast-gp (4 Strains) QC Set, distributed nationwide in the US. The recall was initiated on 18 Feb 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 5112P; Lot Numbers: 5112-01 (Expiration Date 07/31/2020), 5112-02 (Expiration Date 10/31/2020); UDI: 70845357026551
Quantity recalled10 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0587-2022
FDA event ID
89401
Recalling firm
Microbiologics, Saint Cloud, MN
Classification
Class II
Status
Terminated
Recall initiated
18 Feb 2020
FDA classified
8 Feb 2022
Reported
16 Feb 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

18 Feb 2020Recall initiated by company
8 Feb 2022Classified by FDA+721 days
16 Feb 2022Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms

Where it was distributed

Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

Nationwide Puerto Rico

Questions about this recall

Is QC Sets and Panels, Kwik-stik Ast-gp (4 Strains) QC Set recalled?

Yes. Microbiologics recalled QC Sets and Panels, Kwik-stik Ast-gp (4 Strains) QC Set on 18 Feb 2020. The recall is Class II and its status is Terminated. Recall number Z-0587-2022.

Why was it recalled?

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

Which lots are affected?

Catalog Number: 5112P; Lot Numbers: 5112-01 (Expiration Date 07/31/2020), 5112-02 (Expiration Date 10/31/2020); UDI: 70845357026551

Where was it sold?

Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Microbiologics

All 66

Other Test Kits & Diagnostics recalls