CLASS I Device recall · Reported 5 Dec 2018 · FDA Enforcement Report

Quadrox-i Neonatal with integrated arterial filter recalled

Maquet Cardiovascular Us Sales is recalling Quadrox-i Neonatal with integrated arterial filter and Vhk 11000 (Vkmo, distributed nationwide in the US. The recall was initiated on 25 Oct 2018 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI: 04037691773414; Lots: 92209374, 92211752, 92215106, 92222131, 92223169, 92228805, 92234286, 92236876
Quantity recalled490 total in US

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0437-2019
FDA event ID
81382
Recalling firm
Maquet Cardiovascular Us Sales, Wayne, NJ
Classification
Class I
Status
Terminated
Recall initiated
25 Oct 2018
FDA classified
28 Nov 2018
Reported
5 Dec 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

25 Oct 2018Recall initiated by company
28 Nov 2018Classified by FDA+34 days
5 Dec 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Where it was distributed

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.

Nationwide ArizonaCaliforniaFloridaIowaLouisianaNorth CarolinaTexas

Questions about this recall

Why was it recalled?

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Which lots are affected?

UDI: 04037691773414; Lots: 92209374, 92211752, 92215106, 92222131, 92223169, 92228805, 92234286, 92236876

Where was it sold?

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 81382

This product is part of a recall event; the main page for the event is Quadrox-i Neonatal without integrated arterial filter recalled.

More recalls from Maquet Cardiovascular Us Sales

All 62

Other Other Medical Devices recalls

All 13,778