CLASS II Device recall · Reported 26 Aug 2015 · FDA Enforcement Report

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system recalled

Radiometer America is recalling Radiometer Fixation Kit N20 for the TCM CombiM monitoring system. The recall was initiated on 16 Jul 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel #: 905-873 Lots: 407221, 407231, 407241, 407251, 408041, 408191, 408201, 408211, 408231, 408232, 408261, 408271, 409021, 409101, 409231, 409291, 410021, 410141, 410281, 411171, 411201, 411261, 412021, 412111, 412161, 412201, 501081, 501131, 501191, 501311, 502031, 502101, 502131, 502171, 502231, 502261, 503091, 503231, 503311, 504031, 504081, 504111, 504271, 505041, 505061, 505141, 505181
Quantity recalled4244 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2417-2015
FDA event ID
71739
Recalling firm
Radiometer America, Brea, CA
Classification
Class II
Status
Terminated
Recall initiated
16 Jul 2015
FDA classified
19 Aug 2015
Reported
26 Aug 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

16 Jul 2015Recall initiated by company
19 Aug 2015Classified by FDA+34 days
26 Aug 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873. The Fixation Kit N20 is used to apply tc Sensor 84 and tc Sensor 54 on the skin of a patient. TC sensor 84 and 54 are used with the TCM CombiM monitoring system and the Philips TcG10 module. The TCM CombiM monitoring system is intended for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures.

Where it was distributed

Worldwide Distribution - US Distribution and to the countries of : Canada, Spain, France, Great Britain, Philippines, Netherlands, Japan, China, Australia, Denmark, South Africa, Germany, Qatar, Turkey, Hong Kong, Costa Rica, Austria, Finland, Indonesia, Norway, Malaysia, Belgium, Czech Republic, Hungry, United Arab Emirates, Poland, Colombia, Belgium and Malta.

Questions about this recall

Why was it recalled?

When the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Which lots are affected?

Model #: 905-873 Lots: 407221, 407231, 407241, 407251, 408041, 408191, 408201, 408211, 408231, 408232, 408261, 408271, 409021, 409101, 409231, 409291, 410021, 410141, 410281, 411171, 411201, 411261, 412021, 412111, 412161, 412201, 501081, 501131, 501191, 501311, 502031, 502101, 502131, 502171, 502231, 502261, 503091, 503231, 503311, 504031, 504081, 504111, 504271, 505041, 505061, 505141, 505181

Where was it sold?

Worldwide Distribution - US Distribution and to the countries of : Canada, Spain, France, Great Britain, Philippines, Netherlands, Japan, China, Australia, Denmark, South Africa, Germany, Qatar, Turkey, Hong Kong, Costa Rica, Austria, Finland, Indonesia, Norway, Malaysia, Belgium, Czech Republic, Hungry, United Arab Emirates, Poland, Colombia, Belgium and Malta.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.