CLASS II Device recall · Reported 20 Dec 2017 · FDA Enforcement Report
ABL800 analyzer with Flexq module recalled by Radiometer America
Radiometer America is recalling ABL800 analyzer with Flexq module, distributed nationwide in the US. The recall was initiated on 24 Oct 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0231-2018
- FDA event ID
- 78445
- Recalling firm
- Radiometer America, Brea, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 24 Oct 2017
- FDA classified
- 8 Dec 2017
- Reported
- 20 Dec 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Patient Monitors
Timeline
| 24 Oct 2017 | Recall initiated by company | |
| 8 Dec 2017 | Classified by FDA | +45 days |
| 20 Dec 2017 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.
Product as listed by the FDA
ABL800 analyzer with FLEXQ module. Device intended for in vitro testing of samples of whole blood for the parameters pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) - in vitro testing of samples of expired air for the parameters pO2 and pCO2. It is also used for in vitro testing of pleura samples for the pH parameter
Where it was distributed
USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Distributed internationally to Canada.
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Is ABL800 analyzer with Flexq module recalled?
Yes. Radiometer America recalled ABL800 analyzer with Flexq module on 24 Oct 2017. The recall is Class II and its status is Terminated. Recall number Z-0231-2018.
Why was it recalled?
Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.
Which lots are affected?
All serial numbers
Where was it sold?
USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Distributed internationally to Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Radiometer America
All 24| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | ABL90 FLEX Analyzer recalled by Radiometer America | Radiometer America | Device Malfunction | 36 states |
| Device | CLASS II | ABL800 FLEX with Crea recalled by Radiometer America | Radiometer America | Other | 31 states |
| Device | CLASS II | ABL800 analyzer recalled by Radiometer America | Radiometer America | Device Malfunction | Nationwide |
| Device | CLASS II | ABL90 FLEX analyzers recalled by Radiometer America | Radiometer America | Device Malfunction | Nationwide |
| Device | CLASS II | AutoCheck5+, Level 2, specifically designed recalled | Radiometer America | Device Malfunction | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |