CLASS II ONGOING Device recall · Reported 26 Nov 2025 · FDA Enforcement Report
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer recalled
Cellavision AB is recalling REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer, distributed nationwide in the US. The recall was initiated on 8 Oct 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0542-2026
- FDA event ID
- 97835
- Recalling firm
- Cellavision AB, Lund
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Oct 2025
- FDA classified
- 18 Nov 2025
- Reported
- 26 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Test Kits & Diagnostics
Timeline
| 8 Oct 2025 | Recall initiated by company | |
| 18 Nov 2025 | Classified by FDA | +41 days |
| 26 Nov 2025 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Product as listed by the FDA
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.
Where it was distributed
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Which lots are affected?
UDI: 0735004097202/ Serial Numbers 63220-70093 or CN63220-CN70093 Serial Numbers 60001-63219 or CN60001-CN63219
Where was it sold?
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cellavision AB
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CellaVision DM recalled by Cellavision AB | Cellavision AB | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,671| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |
| Device | CLASS II | Techno-Path Multichem AE Unassayed Level recalled | Techno-path Manufacturing | Potency & Assay | Nationwide |
| Device | CLASS II | MAS Omni Immune PRO Liquid Assayed Integrated Immunoassay recalled | Microgenics | Potency & Assay | Nationwide |
| Device | CLASS II | The ABL90 Flex and ABL90 Flex Plus analyzers are intended recalled | Radiometer Medical ApS | Labeling Errors | Nationwide |