CLASS II ONGOING Device recall · Reported 18 Dec 2024 · FDA Enforcement Report

Rover Mobile X-ray System, # MXU-RV19 recalled by Micro-X

Micro-X is recalling Rover Mobile X-ray System, # MXU-RV19, distributed nationwide in the US. The recall was initiated on 23 Oct 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.
Quantity recalled23

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0666-2025
FDA event ID
95593
Recalling firm
Micro-X, Tonsley
Classification
Class II
Status
Ongoing
Recall initiated
23 Oct 2024
FDA classified
12 Dec 2024
Reported
18 Dec 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

23 Oct 2024Recall initiated by company
12 Dec 2024Classified by FDA+50 days
18 Dec 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Product as listed by the FDA

MICRO-X Rover Mobile X-ray System, # MXU-RV19

Where it was distributed

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

Nationwide ArizonaCaliforniaIdahoOhioTexasWashington

Questions about this recall

Why was it recalled?

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Which lots are affected?

UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.

Where was it sold?

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.