CLASS II ONGOING Device recall · Reported 21 Sep 2022 · FDA Enforcement Report

Scenaria View recalled by FUJIFILM Healthcare Americas

FUJIFILM Healthcare Americas is recalling Scenaria View, distributed nationwide in the US. The recall was initiated on 12 Jul 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI (01)04580292766694(11)190926(21)V5001(250)G560045919 (01)04580292766694(11)191002(21)V5002(250)G560045719 (01)04580292766694(11)191008(21)V5004(250)G560045519 (01)04580292766694(11)191107(21)V5005(250)G560046119 (01)04580292766694(11)191113(21)V5006(250)G560048819 (01)04580292766694(11)191218(21)V5007(250)G560052019 (01)04580292766694(11)191220(21)V5008(250)G560052219 (01)04580292766694(11)200124(21)V5009(250)G560052819 (01)04580292766694(11)200124(21)V5010(250)G560052919 (01)04580292766694(11)200224(21)V5011(250)G560053419 (01)04580292766694(11)200514(21)V5017(250)G560053219 (01)04580292766694(11)200904(21)V5018(250)G1Z0169319 (01)04580292766694(11)200911(21)V5019(250)G560055619 (01)04580292766694(11)200915(21)V5020(250)G1Z0169519 (01)04580292766694(11)200925(21)V5021(250)G1Z0170519 (01)04580292766694(11)201001(21)V5022(250)G1Z0170619 (01)04580292766694(11)201009(21)V5023(250)G560059219 (01)04580292766694(11)201014(21)V5024(250)G1Z0000620 (01)04580292766694(11)201020(21)V5026(250)G560077419 (01)04580292766694(11)201020(21)V5027(250)G1Z0001220 (01)04580292766694(11)201029(21)V5028(250)G1Z0001320 (01)04580292766694(11)210108(21)V5031(250)G560001120 (01)04580292766694(11)210112(21)V5032(250)G560001220 (01)04580292766694(11)210115(21)V5033(250)G560001320 (01)04580292766694(11)210208(21)V5034(250)G560002120 (01)04580292766694(11)210216(21)V5035(250)G560002320 (01)04580292766694(11)210217(21)V5036(250)G560002420 (01)04580292766694(11)210222(21)V5037(250)G560002720 (01)04580292766694(11)210226(21)V5038(250)G560001020 (01)04580292766694(11)210410(21)V5039(250)G560003120 (01)04580292766694(11)210414(21)V5040(250)G560003220 (01)04580292766694(11)210518(21)V5041(250)G560003620 (01)04573596200544(11)210713(21)V5042(250)G560007520 (01)04573596200544(11)210713(21)V5043(250)G560003920 (01)04573596200544(11)210729(21)V5044(250)G1Z0233220 (01)04573596200544(11)210729(21)V5045(250)G1Z0245920 (01)04573596200544(11)210902(21)V5046(250)G1Z0127821 (01)04573596200544(11)210903(21)V5047(250)G1Z0127521 (01)04573596200544(11)210909(21)V5048(250)G1Z0128621 (01)04573596200544(11)210922(21)V5049(250)G1Z0128721 (01)04573596200544(11)211004(21)V5050(250)G1Z0130121 (01)04573596200544(11)220302(21)V5051(250)G1Z0208921 (01)04573596200544(11)220304(21)V5052(250)G1Z0209021 (01)04573596200544(11)220307(21)V5053(250)G1Z0214121 (01)04573596200544(11)220330(21)V5054(250)G1Z0236821 Serial Numbers: G560045919 G560045719 G560045519 G560046119 G560048819 G560052019 G560052219 G560052819 G560052919 G560053419 G560053219 G1Z0169319 G560055619 G1Z0169519 G1Z0170519 G1Z0170619 G560059219 G1Z0000620 G560077419 G1Z0001220 G1Z0001320 G560001120 G560001220 G560001320 G560002120 G560002320 G560002420 G560002720 G560001020 G560003120 G560003220 G560003620 G560007520 G560003920 G1Z0233220 G1Z0245920 G1Z0127821 G1Z0127521 G1Z0128621 G1Z0128721 G1Z0130121 G1Z0208921 G1Z0209021 G1Z0214121 G1Z0236821
Quantity recalled46 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1748-2022
FDA event ID
90751
Recalling firm
FUJIFILM Healthcare Americas, Lexington, MA
Classification
Class II
Status
Ongoing
Recall initiated
12 Jul 2022
FDA classified
15 Sep 2022
Reported
21 Sep 2022
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction
Device type
Medical Software

Timeline

12 Jul 2022Recall initiated by company
15 Sep 2022Classified by FDA+65 days
21 Sep 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

SCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Is Scenaria View recalled?

Yes. FUJIFILM Healthcare Americas recalled Scenaria View on 12 Jul 2022. The recall is Class II and its status is Ongoing. Recall number Z-1748-2022.

Why was it recalled?

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

Which lots are affected?

UDI (01)04580292766694(11)190926(21)V5001(250)G560045919 (01)04580292766694(11)191002(21)V5002(250)G560045719 (01)04580292766694(11)191008(21)V5004(250)G560045519 (01)04580292766694(11)191107(21)V5005(250)G560046119 (01)04580292766694(11)191113(21)V5006(250)G560048819 (01)04580292766694(11)191218(21)V5007(250)G560052019 (01)04580292766694(11)191220(21)V5008(250)G560052219 (01)04580292766694(11)200124(21)V5009(250)G560052819 (01)04580292766694(11)200124(21)V5010(250)G560052919 (01)04580292766694(11)200224(21)V5011(250)G560053419 (01)04580292766694(11)200514(21)V5017(250)G560053219 (01)04580292766694(11)200904(21)V5018(250)G1Z0169319 (01)04580292766694(11)200911(21)V5019(250)G560055619 (01)04580292766694(11)200915(21)V5020(250)G1Z0169519 (01)04580292766694(11)200925(21)V5021(250)G1Z0170519 (01)04580292766694(11)201001(21)V5022(250)G1Z0170619 (01)04580292766694(11)201009(21)V5023(250)G560059219 (01)04580292766694(11)201014(21)V5024(250)G1Z0000620 (01)04580292766694(11)201020(21)V5026(250)G560077419 (01)04580292766694(11)201020(21)V5027(250)G1Z0001220 (01)04580292766694(11)201029(21)V5028(250)G1Z0001320 (01)04580292766694(11)210108(21)V5031(250)G560001120 (01)04580292766694(11)210112(21)V5032(250)G560001220 (01)04580292766694(11)210115(21)V5033(250)G560001320 (01)04580292766694(11)210208(21)V5034(250)G560002120 (01)04580292766694(11)210216(21)V5035(250)G560002320 (01)04580292766694(11)210217(21)V5036(250)G560002420 (01)04580292766694(11)210222(21)V5037(250)G560002720 (01)04580292766694(11)210226(21)V5038(250)G560001020 (01)04580292766694(11)210410(21)V5039(250)G560003120 (01)04580292766694(11)210414(21)V5040(250)G560003220 (01)04580292766694(11)210518(21)V5041(250)G560003620 (01)04573596200544(11)210713(21)V5042(250)G560007520 (01)04573596200544(11)210713(21)V5043(250)G560003920 (01)04573596200544(11)210729(21)V5044(250)G1Z0233220 (01)04573596200544(11)210729(21)V5045(250)G1Z0245920 (01)04573596200544(11)210902(21)V5046(250)G1Z0127821 (01)04573596200544(11)210903(21)V5047(250)G1Z0127521 (01)04573596200544(11)210909(21)V5048(250)G1Z0128621 (01)04573596200544(11)210922(21)V5049(250)G1Z0128721 (01)04573596200544(11)211004(21)V5050(250)G1Z0130121 (01)04573596200544(11)220302(21)V5051(250)G1Z0208921 (01)04573596200544(11)220304(21)V5052(250)G1Z0209021 (01)04573596200544(11)220307(21)V5053(250)G1Z0214121 (01)04573596200544(11)220330(21)V5054(250)G1Z0236821 Serial Numbers: G560045919 G560045719 G560045519 G560046119 G560048819 G560052019 G560052219 G560052819 G560052919 G560053419 G560053219 G1Z0169319 G560055619 G1Z0169519 G1Z0170519 G1Z0170619 G560059219 G1Z0000620 G560077419 G1Z0001220 G1Z0001320 G560001120 G560001220 G560001320 G560002120 G560002320 G560002420 G560002720 G560001020 G560003120 G560003220 G560003620 G560007520 G560003920 G1Z0233220 G1Z0245920 G1Z0127821 G1Z0127521 G1Z0128621 G1Z0128721 G1Z0130121 G1Z0208921 G1Z0209021 G1Z0214121 G1Z0236821

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.