CLASS II ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report
Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis recalled
Hologic is recalling Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis, distributed nationwide in the US. The recall was initiated on 19 Aug 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3248-2024
- FDA event ID
- 95286
- Recalling firm
- Hologic, Newark, DE
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 19 Aug 2024
- FDA classified
- 25 Sep 2024
- Reported
- 2 Oct 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 19 Aug 2024 | Recall initiated by company | |
| 25 Sep 2024 | Classified by FDA | +37 days |
| 2 Oct 2024 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.1 SYSTEMS 3. ASY-17219 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.10.0 SYSTEMS 4. ASY-17220 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.1.0 SYSTEMS 5. ASY-17221 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.0 SYSTEMS 6. ASY-17222 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.0 SYSTEMS 7. 3DM-LIC-SW-UP-V2.3 DIMENSIONS SOFTWARE 2.3 (3DIMENSIONS) 8. SDM-LIC-SW-UP-V1.12 DIMENSION SOFTWARE 1.12 (SELENIA DIMENSIONS)
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands.
Questions about this recall
Why was it recalled?
Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis
Which lots are affected?
UDI-DI: 15420045505636, 15420045511897 Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View¿ software
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hologic
All 48| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | 3dimensions Mammography System recalled by Hologic | Hologic | Other | Nationwide |
| Device | CLASS II | Selenia Dimensions Mammography Systems recalled by Hologic | Hologic | Other | Nationwide |
| Device | CLASS II | Genius" Review Station Display (Barco Monitor) recalled by Hologic | Hologic | Device Malfunction | Nationwide |
| Device | CLASS II | Aptima HPV Assay recalled by Hologic | Hologic | Potency & Assay | Nationwide |
| Device | CLASS II | Brevera Breast Biopsy System Disposable Needle recalled by Hologic | Hologic | Particulate Matter | Nationwide |
Other Medical Software recalls
All 935| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Metatarsophalangeal (MTP) Implant is a sterile recalled by restor3d | restor3d | Sterility | Nationwide |
| Device | CLASS II | Ion Torrent Genexus Dx recalled by Life Technologies Holdings Pte | Life Technologies Holdings Pte | Device Malfunction | Nationwide |
| Device | CLASS II | Mulitple-type osteoarticular pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |